Philips M5066A owner manual Patient treatment protocol options

Page 52

F-2

Patient treatment protocol options

parameter

“call EMS” voice reminder timing

shock series

shock series interval (minutes)

settings

default

default description

• At power on (when

At the start of

Provides a voice reminder to make

the user turns on

the first

sure emergency medical services have

the HeartStart)

patient care

been called, at the start of the first

• At power on and at

pause

patient care pause.

 

 

the start of the first

 

 

patient care pause

 

 

• At the start of the

 

 

first patient care

 

 

pause

 

 

• No reminder

 

 

1, 2, 3, 4

1

The automatic protocol pause for CPR

 

 

is activated each time a shock is

 

 

delivered.*

 

 

During the protocol pause, the

 

 

HeartStart does not perform rhythm

 

 

analysis.

 

 

The length of the protocol pause after

 

 

a shock series is completed is

 

 

determined by the protocol pause

 

 

timer setting.

1.0, 2.0,

1.0

A delivered shock must occur within 1

(infinity)

 

minute of the previous shock to be

 

 

counted as part of the current shock

 

 

series.

 

 

NOTE: This parameter is only applicable

 

 

when the shock series is not configured to

 

 

the default 1 shock.

Philips Medical Systems

*A shock series begins when a shock is delivered after the HeartStart is turned on. A new shock series begins after a protocol pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous shock exceeds the shock series interval setting.

HEARTSTART M5066A

Image 52
Contents M5066A EditionIntentionally blank HeartStart Defibrillator M5066A Pads Cartridge HandleOn/Off Button. Press this Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank About this edition Edition historyCopyright Authorized EU representativeIntentionally blank Contents Appendices Introduction to the HeartStart DescriptionSudden Cardiac Arrest Indications for UseNational and local requirements Training and practiceFor more information Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A After each use After using the HeartStartHeartStart data storage Heartstart M5066A Maintaining the HeartStart Routine MaintenancePeriodic checks Troubleshooting tips Cleaning the HeartStartDisposing of the HeartStart Accessories for the HeartStart REF 68-PCHATHeartstart M5066A Glossary of terms AEDCPR ECGNSA Smart NSA pause SCA Symbol Description Glossary of symbols/controlsDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Precautions Contact with the patientRegularly for damage, according to directions Metal objects such as a bed frame or stretcherHeartStart Defibrillator specifications Technical informationPhysical EnvironmentalControls and indicators Category SealingDefibrillation Waveform Category Waveform parametersDelivered Current a Energy JShock-to-shock cycle time Disarm AED modeNewborn 14 J/kg Category Function Shockable rhythms Non-shockable rhythms ECG analysis systemAnalysis protocol ECG analysis performance Accessories specifications Battery M5070A Category Specifications Battery type Prior to insertion Standby lifeWhole and may also endanger human health Environmental considerationsRegulations Configuration Parameter Speaker volumeDevice options ECG out dataPatient treatment protocol options NSA pause type Settings Standard NSADuring a Smart NSA pause, Pause HeartStart Monitoring duringCPR4 Installed Infants and children when an infantChild pads cartridge is installed Yes, NoHeartstart M5066A Testing Testing and troubleshootingTroubleshooting Green Ready light is solid HeartStart tells you Troubleshooting while the HeartStart is in useTo insert a pads cartridge HeartStart tells you To stop all motion Contact Philips for service if needed Troubleshooting while the HeartStart is not in useGreen Ready light is not on Behavior Electromagnetic conformity Electromagnetic emissionsEmissions test CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901