Philips M5066A owner manual Maintaining the HeartStart, Routine Maintenance, Periodic checks

Page 29

Philips Medical Systems

5 Maintaining the HeartStart

Routine Maintenance

The HeartStart is very simple to maintain. The defibrillator performs a self-test every day. In addition, a battery insertion self-test is run whenever a battery is installed in the device. The defibrillator’s extensive automatic self-test features eliminate the need for any manual calibration. The HeartStart has no user- serviceable parts.

WARNING: Electrical shock hazard. Do not open the HeartStart, remove its covers, or attempt repair. There are no user-serviceable components in the HeartStart. If repair is required, return the HeartStart to Philips for service.

Reminders:

Do not leave the HeartStart without a pads cartridge installed; the defibrillator will start chirping and the i-button will start flashing. For directions on changing the pads cartridge, see Chapter 2, “Setting up the

HeartStart.”

5

The HeartStart runs daily self-tests. As long as the green Ready light is blinking, it is not necessary to test the defibrillator by initiating a battery insertion self-test. This uses battery power and risks draining the battery prematurely.

Periodic checks

Other than the checks recommended after each use of the HeartStart, maintenance is limited to periodically checking the following:

Check the green Ready light. If the green Ready light is not blinking, see Troubleshooting Tips, below.

Replace any used, damaged or expired supplies and accessories

Check the outside of the defibrillator. If you see cracks or other signs of damage, contact Philips for technical support.

5-1

Image 29
Contents Edition M5066AIntentionally blank Pads Cartridge Handle HeartStart Defibrillator M5066AOn/Off Button. Press this Information Button. ThisIntentionally blank Quick Reference Intentionally blank Important Note HeartStartIntentionally blank Edition history About this editionCopyright Authorized EU representativeIntentionally blank Contents Appendices Description Introduction to the HeartStartSudden Cardiac Arrest Indications for UseFor more information Training and practiceNational and local requirements Package contents Setting up the HeartStartHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Overview Using the HeartStartPull the green handle Place the pads If a shock is needed Press the Shock buttonRemove the Infant/Child Pads Cartridge from its package Treating infants and childrenDo not Delay Treatment When emergency medical services arrivePhilips Medical Systems Heartstart M5066A HeartStart data storage After using the HeartStartAfter each use Heartstart M5066A Periodic checks Routine MaintenanceMaintaining the HeartStart Disposing of the HeartStart Cleaning the HeartStartTroubleshooting tips REF 68-PCHAT Accessories for the HeartStartHeartstart M5066A AED Glossary of termsCPR ECGNSA Smart NSA pause SCA Defibrillator and opens pads cartridge for use Glossary of symbols/controlsSymbol Description Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Contact with the patient PrecautionsRegularly for damage, according to directions Metal objects such as a bed frame or stretcherTechnical information HeartStart Defibrillator specificationsPhysical EnvironmentalCategory Sealing Controls and indicatorsCategory Waveform parameters Defibrillation WaveformDelivered Current a Energy JNewborn 14 J/kg Disarm AED modeShock-to-shock cycle time Analysis protocol ECG analysis systemCategory Function Shockable rhythms Non-shockable rhythms ECG analysis performance Battery M5070A Accessories specificationsCategory Specifications Battery type Prior to insertion Standby lifeRegulations Environmental considerationsWhole and may also endanger human health Parameter Speaker volume ConfigurationDevice options ECG out dataPatient treatment protocol options Standard NSA NSA pause type SettingsDuring a Smart NSA pause, Pause HeartStart Monitoring duringCPR4 Infants and children when an infant InstalledChild pads cartridge is installed Yes, NoHeartstart M5066A Troubleshooting Testing and troubleshootingTesting To insert a pads cartridge Troubleshooting while the HeartStart is in useGreen Ready light is solid HeartStart tells you HeartStart tells you To stop all motion Green Ready light is not on Behavior Troubleshooting while the HeartStart is not in useContact Philips for service if needed Electromagnetic emissions Electromagnetic conformityEmissions test CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Environmental considerations Important warnings and remindersIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901