Philips M5066A owner manual Glossary of terms, Aed, Cpr, Ecg

Page 33

Philips Medical Systems

B

Glossary of terms

 

 

 

B

 

The terms listed in this Glossary are defined in the context of the Philips

 

 

 

HeartStart Defibrillator and its use.

 

AED

Automated external defibrillator (a semi-automatic defibrillator).

 

AED mode

The standard treatment mode for the HeartStart Defibrillator. It provides voice

 

 

instructions guiding the rescuer through applying the adhesive pads, waiting for

 

 

rhythm analysis, and delivering a shock if needed.

 

analysis

See “SMART analysis.”

 

arrhythmia

An unhealthy, often irregular, beating of the heart.

 

artifact

Electrical “noise” caused by sources such as muscle movements, CPR, patient

 

 

transport, or static electricity that may interfere with rhythm analysis.

 

battery

The sealed lithium manganese dioxide battery used to power the HeartStart

 

 

Defibrillator. It is provided in a pack that fits into a compartment on the back of

 

 

the defibrillator.

 

Caution light

A triangular light on the front of the HeartStart Defibrillator that flashes during

 

 

rhythm analysis and is on solid when a shock is advised, as a reminder not to touch

 

 

the patient.

 

configuration

The settings for all operating options of the HeartStart Defibrillator, including

 

 

treatment protocol. The factory default configuration can be modified by

 

 

authorized personnel using HeartStart Event Review software.

 

CPR

Cardiopulmonary resuscitation. A technique for providing artificial respiration and

 

 

heart compressions.

 

CPR Coaching

Basic verbal instructions for performing cardiopulmonary resuscitation, including

 

 

hand placement, rescue breathing, compression depth and timing, provided by the

 

 

HeartStart when the flashing blue i-button is pressed during the first 30 seconds of

 

 

a patient care pause.

 

defibrillation

Termination of cardiac fibrillation by applying electrical energy.

 

ECG

Electrocardiogram, a record of the electrical rhythm of the heart as detected

 

 

through defibrillation pads.

 

B-1

Image 33
Contents Edition M5066AIntentionally blank Pads Cartridge Handle HeartStart Defibrillator M5066AOn/Off Button. Press this Information Button. ThisIntentionally blank Quick Reference Intentionally blank Important Note HeartStartIntentionally blank Edition history About this editionCopyright Authorized EU representativeIntentionally blank Contents Appendices Description Introduction to the HeartStartSudden Cardiac Arrest Indications for UseTraining and practice National and local requirementsFor more information Package contents Setting up the HeartStartHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Overview Using the HeartStartPull the green handle Place the pads If a shock is needed Press the Shock buttonRemove the Infant/Child Pads Cartridge from its package Treating infants and childrenDo not Delay Treatment When emergency medical services arrivePhilips Medical Systems Heartstart M5066A After using the HeartStart After each useHeartStart data storage Heartstart M5066A Routine Maintenance Maintaining the HeartStartPeriodic checks Cleaning the HeartStart Troubleshooting tipsDisposing of the HeartStart REF 68-PCHAT Accessories for the HeartStartHeartstart M5066A AED Glossary of termsCPR ECGNSA Smart NSA pause SCA Glossary of symbols/controls Symbol DescriptionDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Contact with the patient PrecautionsRegularly for damage, according to directions Metal objects such as a bed frame or stretcherTechnical information HeartStart Defibrillator specificationsPhysical EnvironmentalCategory Sealing Controls and indicatorsCategory Waveform parameters Defibrillation WaveformDelivered Current a Energy JDisarm AED mode Shock-to-shock cycle timeNewborn 14 J/kg ECG analysis system Category Function Shockable rhythms Non-shockable rhythmsAnalysis protocol ECG analysis performance Battery M5070A Accessories specificationsCategory Specifications Battery type Prior to insertion Standby lifeEnvironmental considerations Whole and may also endanger human healthRegulations Parameter Speaker volume ConfigurationDevice options ECG out dataPatient treatment protocol options Standard NSA NSA pause type SettingsDuring a Smart NSA pause, Pause HeartStart Monitoring duringCPR4 Infants and children when an infant InstalledChild pads cartridge is installed Yes, NoHeartstart M5066A Testing and troubleshooting TestingTroubleshooting Troubleshooting while the HeartStart is in use Green Ready light is solid HeartStart tells youTo insert a pads cartridge HeartStart tells you To stop all motion Troubleshooting while the HeartStart is not in use Contact Philips for service if neededGreen Ready light is not on Behavior Electromagnetic emissions Electromagnetic conformityEmissions test CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Environmental considerations Important warnings and remindersIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901
Related manuals
Manual 68 pages 2.08 Kb

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.