Philips M5066A Yes, No, Optional CPR Coaching includes, By the CPR Coaching, Installed, Adult

Page 55

Philips Medical Systems

parameter

CPR Coaching adult ventilation instruction

CPR Coaching infant/child ventilation instruction

CPR Coaching compression:ventilation ratio

F-5

settings

default

default description

Yes, No

Yes

Optional CPR Coaching includes

 

 

rescue breaths at the rate determined

 

 

by the CPR Coaching

 

 

compression:ventilation ratio for

 

 

adults when an adult pads cartridge is

 

 

installed.

 

 

NOTE: if this parameter is configured to

 

 

NO, CPR Coaching will always be

 

 

compressions-only when an adult pads

 

 

cartridge is installed.

Yes, No

Yes

Optional CPR Coaching includes

 

 

rescue breaths at the rate determined

 

 

by the CPR Coaching

 

 

compression:ventilation ratio for

 

 

infants and children when an infant/

 

 

child pads cartridge is installed.

 

 

NOTE: if this parameter is configured to

 

 

NO, CPR Coaching will always be

 

 

compressions-only when an infant/child

 

 

pads cartridge is installed.

• 30:2 adult and

30:2 adult and

If the user presses the i-button for

30:2 infant/child

30:2 infant/

optional CPR Coaching during a

• 30:2 adult and

child

protocol pause or NSA pause, the

 

HeartStart provides coaching in basic

15:2 infant/child

 

 

CPR for cycles of 30 compressions and

• 15:2 adult and

 

 

2 ventilations for adults, children, and

15:2 infant/child

 

infants. Pauses begin and end with

 

 

compressions.

F

Image 55
Contents Edition M5066AIntentionally blank Information Button. This HeartStart Defibrillator M5066APads Cartridge Handle On/Off Button. Press thisIntentionally blank Quick Reference Intentionally blank Important Note HeartStartIntentionally blank Authorized EU representative About this editionEdition history CopyrightIntentionally blank Contents Appendices Indications for Use Introduction to the HeartStartDescription Sudden Cardiac ArrestNational and local requirements Training and practiceFor more information Package contents Setting up the HeartStartHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Overview Using the HeartStartPull the green handle Place the pads If a shock is needed Press the Shock buttonRemove the Infant/Child Pads Cartridge from its package Treating infants and childrenDo not Delay Treatment When emergency medical services arrivePhilips Medical Systems Heartstart M5066A After each use After using the HeartStartHeartStart data storage Heartstart M5066A Maintaining the HeartStart Routine MaintenancePeriodic checks Troubleshooting tips Cleaning the HeartStartDisposing of the HeartStart REF 68-PCHAT Accessories for the HeartStartHeartstart M5066A ECG Glossary of termsAED CPRNSA Smart NSA pause SCA Symbol Description Glossary of symbols/controlsDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Metal objects such as a bed frame or stretcher PrecautionsContact with the patient Regularly for damage, according to directionsEnvironmental HeartStart Defibrillator specificationsTechnical information PhysicalCategory Sealing Controls and indicatorsCurrent a Energy J Defibrillation WaveformCategory Waveform parameters DeliveredShock-to-shock cycle time Disarm AED modeNewborn 14 J/kg Category Function Shockable rhythms Non-shockable rhythms ECG analysis systemAnalysis protocol ECG analysis performance Prior to insertion Standby life Accessories specificationsBattery M5070A Category Specifications Battery typeWhole and may also endanger human health Environmental considerationsRegulations ECG out data ConfigurationParameter Speaker volume Device options Patient treatment protocol options Monitoring during NSA pause type SettingsStandard NSA During a Smart NSA pause, Pause HeartStartCPR4 Yes, No InstalledInfants and children when an infant Child pads cartridge is installedHeartstart M5066A Testing Testing and troubleshootingTroubleshooting Green Ready light is solid HeartStart tells you Troubleshooting while the HeartStart is in useTo insert a pads cartridge HeartStart tells you To stop all motion Contact Philips for service if needed Troubleshooting while the HeartStart is not in useGreen Ready light is not on Behavior Cispr Electromagnetic conformityElectromagnetic emissions Emissions testElectromagnetic immunity Separation distance according to frequency of transmitter m Environmental considerations Important warnings and remindersIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901