Philips M5066A owner manual CPR4

Page 54

F-4

parameter

NSA pause timer (minutes)

CPR prompt

settings

0.5, 1.0, 1.5, 2.0, 2.5, 3.0

CPR1: Instructs the user to begin CPR.

CPR2: Instructs the user that it is safe to touch the patient and to begin CPR.

CPR3: Instructs the user to begin CPR and to press the i-button for CPR Coaching.

CPR4: Instructs the user that it is safe to touch the patient, to begin CPR, and to press the i-button for CPR Coaching.

default

default description

2.0A 2-minute NSA pause for CPR automatically starts after voice

instruction is given when no shock is advised (NSA).*

If the user presses the i-button for optional CPR coaching, the HeartStart provides coaching for 5 cycles of CPR, starting and ending with compressions, when the CPR Coaching parameters are also set to their default values. The number of CPR cycles varies for other NSA pause timer and CPR Coaching parameter settings.

Note: Because the NSA pause ends upon completion of a CPR cycle in order to maximize the benefits of CPR, the actual duration of the pause may differ slightly from the timer setting.

CPR4:

The CPR reminder voice instructions

Instructs the

provided at the beginning of a pause

user that it is

interval assures the user that it is safe

safe to touch

to touch the patient, instructs the user

the patient, to

to begin CPR, and invites the user to

begin CPR, and

press the i-button for guidance in the

to press the

basic steps of CPR.

i-button for

Note: CPR Coaching is available only with

CPR Coaching.

the CPR3 and CPR4 settings.

 

Philips Medical Systems

*If the shock series is configured to 2 or more, and a shock has been delivered as part of a series, the length of the first NSA pause within that shock series is determined by the protocol pause timer setting. Otherwise, the length of an NSA pause is determined by the NSA pause timer setting.

HEARTSTART M5066A

Image 54
Contents M5066A EditionIntentionally blank On/Off Button. Press this HeartStart Defibrillator M5066APads Cartridge Handle Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank Copyright About this editionEdition history Authorized EU representativeIntentionally blank Contents Appendices Sudden Cardiac Arrest Introduction to the HeartStartDescription Indications for UseTraining and practice National and local requirementsFor more information Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A After using the HeartStart After each useHeartStart data storage Heartstart M5066A Routine Maintenance Maintaining the HeartStartPeriodic checks Cleaning the HeartStart Troubleshooting tipsDisposing of the HeartStart Accessories for the HeartStart REF 68-PCHATHeartstart M5066A CPR Glossary of termsAED ECGNSA Smart NSA pause SCA Glossary of symbols/controls Symbol DescriptionDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Regularly for damage, according to directions PrecautionsContact with the patient Metal objects such as a bed frame or stretcherPhysical HeartStart Defibrillator specificationsTechnical information EnvironmentalControls and indicators Category SealingDelivered Defibrillation WaveformCategory Waveform parameters Current a Energy JDisarm AED mode Shock-to-shock cycle timeNewborn 14 J/kg ECG analysis system Category Function Shockable rhythms Non-shockable rhythmsAnalysis protocol ECG analysis performance Category Specifications Battery type Accessories specificationsBattery M5070A Prior to insertion Standby lifeEnvironmental considerations Whole and may also endanger human healthRegulations Device options Configuration Parameter Speaker volume ECG out dataPatient treatment protocol options During a Smart NSA pause, Pause HeartStart NSA pause type SettingsStandard NSA Monitoring duringCPR4 Child pads cartridge is installed InstalledInfants and children when an infant Yes, NoHeartstart M5066A Testing and troubleshooting TestingTroubleshooting Troubleshooting while the HeartStart is in use Green Ready light is solid HeartStart tells youTo insert a pads cartridge HeartStart tells you To stop all motion Troubleshooting while the HeartStart is not in use Contact Philips for service if neededGreen Ready light is not on Behavior Emissions test Electromagnetic conformityElectromagnetic emissions CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901