Philips M5066A owner manual ECG analysis performance

Page 48

E-6

ECG analysis performance

rhythm class

shockable rhythm — ventricular fibrillation

shockable rhythm — ventricular tachycardia

non-shockable rhythm — normal sinus rhythm

non-shockable rhythm — asystole

non-shockable rhythm — all other non-shockable rhythmsc

ECG test

meets AHA recommendationsb for adult defibrillation

observed

90% one-sided

sample

a

size

 

performance

lower confidence limit

 

 

 

300

 

sensitivity >90%

(87%)

 

 

 

(meets AAMI DF80 requirement)

100

 

sensitivity >75%

(67%)

 

 

 

(meets AAMI DF80 requirement)

300

 

specificity >99%

(97%)

 

 

 

(meets AAMI DF80 requirement)

100

 

specificity >95%

(92%)

 

 

 

(meets AAMI DF80 requirement)

450

 

specificity >95%

(88%)

 

 

 

(meets AAMI DF80 requirement)

Philips

a.From Philips Medical Systems Heartstream ECG rhythm databases.

b.American Heart Association (AHA) AED Task Force, Subcommittee on AED Safety & Efficacy. Automatic External Defibrillators for Public Access Use: Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporation of New Waveforms, and Enhancing Safety. Circulation 1997;95:1677-1682.

c.Supraventricular tachycardia (SVT) is specifically included in the non-shockable rhythm class, in accordance with AHA recommendationsb and the AAMI standard DF80.

Medical Systems

HEARTSTART M5066A

Image 48
Contents M5066A EditionIntentionally blank HeartStart Defibrillator M5066A Pads Cartridge HandleOn/Off Button. Press this Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank About this edition Edition historyCopyright Authorized EU representativeIntentionally blank Contents Appendices Introduction to the HeartStart DescriptionSudden Cardiac Arrest Indications for UseTraining and practice National and local requirementsFor more information Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A After using the HeartStart After each useHeartStart data storage Heartstart M5066A Routine Maintenance Maintaining the HeartStartPeriodic checks Cleaning the HeartStart Troubleshooting tipsDisposing of the HeartStart Accessories for the HeartStart REF 68-PCHATHeartstart M5066A Glossary of terms AEDCPR ECGNSA Smart NSA pause SCA Glossary of symbols/controls Symbol DescriptionDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Precautions Contact with the patientRegularly for damage, according to directions Metal objects such as a bed frame or stretcherHeartStart Defibrillator specifications Technical informationPhysical EnvironmentalControls and indicators Category Sealing Defibrillation Waveform Category Waveform parameters Delivered Current a Energy JDisarm AED mode Shock-to-shock cycle timeNewborn 14 J/kg ECG analysis system Category Function Shockable rhythms Non-shockable rhythmsAnalysis protocol ECG analysis performance Accessories specifications Battery M5070ACategory Specifications Battery type Prior to insertion Standby lifeEnvironmental considerations Whole and may also endanger human healthRegulations Configuration Parameter Speaker volumeDevice options ECG out dataPatient treatment protocol options NSA pause type Settings Standard NSADuring a Smart NSA pause, Pause HeartStart Monitoring duringCPR4 Installed Infants and children when an infantChild pads cartridge is installed Yes, NoHeartstart M5066A Testing and troubleshooting TestingTroubleshooting Troubleshooting while the HeartStart is in use Green Ready light is solid HeartStart tells youTo insert a pads cartridge HeartStart tells you To stop all motion Troubleshooting while the HeartStart is not in use Contact Philips for service if neededGreen Ready light is not on Behavior Electromagnetic conformity Electromagnetic emissionsEmissions test CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901