HeartStart tells you:

... to stop all motion

... the shock was not delivered

... the shock button was not pressed

possible cause

The patient is being moved or jostled.

The environment is dry and movement around the patient is causing static electricity to interfere with ECG analysis.

Radio or electrical sources are interfering with ECG analysis.

The pads may not be making good contact with the patient’s skin.

The pads may be touching each other.

The pads may be damaged.

Shock has been advised but the shock button has not been pressed within 30 seconds.

G-3

recommended action

Stop CPR; do not touch the patient. Minimize patient motion. If the patient is being transported, stop the vehicle.

Responders and bystanders should minimize motion, particularly in dry environments that can generate static electricity.

Check for possible causes of radio and electrical interference and turn them off or remove them from the area.

Press the pads firmly to the patient's chest.

Make sure the adhesive pads are correctly positioned on the patient.

Replace the pads if necessary.

When next prompted, press the Shock button to deliver shock.

G

Image 59
Contents Edition M5066AIntentionally blank Information Button. This HeartStart Defibrillator M5066APads Cartridge Handle On/Off Button. Press thisIntentionally blank Quick Reference Intentionally blank Important Note HeartStartIntentionally blank Authorized EU representative About this editionEdition history CopyrightIntentionally blank Contents Appendices Indications for Use Introduction to the HeartStartDescription Sudden Cardiac ArrestFor more information Training and practiceNational and local requirements Package contents Setting up the HeartStartHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Overview Using the HeartStartPull the green handle Place the pads If a shock is needed Press the Shock buttonRemove the Infant/Child Pads Cartridge from its package Treating infants and childrenDo not Delay Treatment When emergency medical services arrivePhilips Medical Systems Heartstart M5066A HeartStart data storage After using the HeartStartAfter each use Heartstart M5066A Periodic checks Routine MaintenanceMaintaining the HeartStart Disposing of the HeartStart Cleaning the HeartStartTroubleshooting tips REF 68-PCHAT Accessories for the HeartStartHeartstart M5066A ECG Glossary of termsAED CPRNSA Smart NSA pause SCA Defibrillator and opens pads cartridge for use Glossary of symbols/controlsSymbol Description Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Metal objects such as a bed frame or stretcher PrecautionsContact with the patient Regularly for damage, according to directionsEnvironmental HeartStart Defibrillator specificationsTechnical information PhysicalCategory Sealing Controls and indicatorsCurrent a Energy J Defibrillation WaveformCategory Waveform parameters DeliveredNewborn 14 J/kg Disarm AED modeShock-to-shock cycle time Analysis protocol ECG analysis systemCategory Function Shockable rhythms Non-shockable rhythms ECG analysis performance Prior to insertion Standby life Accessories specificationsBattery M5070A Category Specifications Battery typeRegulations Environmental considerationsWhole and may also endanger human health ECG out data ConfigurationParameter Speaker volume Device optionsPatient treatment protocol options Monitoring during NSA pause type SettingsStandard NSA During a Smart NSA pause, Pause HeartStartCPR4 Yes, No InstalledInfants and children when an infant Child pads cartridge is installedHeartstart M5066A Troubleshooting Testing and troubleshootingTesting To insert a pads cartridge Troubleshooting while the HeartStart is in useGreen Ready light is solid HeartStart tells you HeartStart tells you To stop all motion Green Ready light is not on Behavior Troubleshooting while the HeartStart is not in useContact Philips for service if needed Cispr Electromagnetic conformityElectromagnetic emissions Emissions testElectromagnetic immunity Separation distance according to frequency of transmitter m Environmental considerations Important warnings and remindersIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901
Related manuals
Manual 68 pages 2.08 Kb

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.