Philips M5066A owner manual Shock-to-shock cycle time, Disarm AED mode, Newborn 14 J/kg

Page 46

E-4

category

energy* (pediatric doses indicated are based on CDC growth charts for the 50th percentile weights for boys.)

charge control

“charge complete” indicator

shock-to-shock cycle time

patient care pause-to-shock time

disarm (AED mode)

adult shock delivery vector

infant/child shock delivery vector

specifications

Using HeartStart Adult SMART Pads: 150 J nominal (±15%) into a 50 ohm load.

Using HeartStart Infant/Child SMART Pads: 50 J nominal (±15%) into a 50 ohm load. Sample pediatric energy doses:

age

energy dose

newborn

14 J/kg

1 year

5 J/kg

2 − 3 years

4 J/kg

4 − 5 years

3 J/kg

6 − 8 years

2 J/kg

*National Center for Health Statistics in collaboration with the National Center for Chronic Disease Prevention and Health Promotion. CDC growth charts: weight-for-age percentiles, revised and corrected November 28, 2000. Atlanta, GA: Centers for Disease Control and Prevention © 2000.

Controlled by Patient Analysis System for automated operation.

Shock button flashes, audio tone sounds.

<20 seconds, typical, including analysis.

Quick Shock. 8 seconds, typical, from end of patient care pause to shock delivery.

Once charged, the defibrillator will disarm if:

the patient’s heart rhythm changes to non-shockable rhythm,

a shock is not delivered within 30 seconds after the defibrillator has charged for shock delivery,

the On/Off button is pressed and held down for at least one (1) second to turn off the defibrillator,

the adhesive pads are removed from the patient or the pads cartridge is disconnected from the defibrillator,

the battery is removed or is completely depleted or

the impedance between pads is out of range.

Via adhesive pads placed in the anterior-anterior (Lead II) position.

Via adhesive pads typically placed in the anterior-posterior position.

Philips Medical Systems

HEARTSTART M5066A

Image 46
Contents M5066A EditionIntentionally blank On/Off Button. Press this HeartStart Defibrillator M5066APads Cartridge Handle Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank Copyright About this editionEdition history Authorized EU representativeIntentionally blank Contents Appendices Sudden Cardiac Arrest Introduction to the HeartStartDescription Indications for UseNational and local requirements Training and practiceFor more information Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A After each use After using the HeartStartHeartStart data storage Heartstart M5066A Maintaining the HeartStart Routine MaintenancePeriodic checks Troubleshooting tips Cleaning the HeartStartDisposing of the HeartStart Accessories for the HeartStart REF 68-PCHATHeartstart M5066A CPR Glossary of termsAED ECGNSA Smart NSA pause SCA Symbol Description Glossary of symbols/controlsDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Regularly for damage, according to directions PrecautionsContact with the patient Metal objects such as a bed frame or stretcher Physical HeartStart Defibrillator specifications Technical information EnvironmentalControls and indicators Category SealingDelivered Defibrillation WaveformCategory Waveform parameters Current a Energy JShock-to-shock cycle time Disarm AED modeNewborn 14 J/kg Category Function Shockable rhythms Non-shockable rhythms ECG analysis systemAnalysis protocol ECG analysis performance Category Specifications Battery type Accessories specificationsBattery M5070A Prior to insertion Standby lifeWhole and may also endanger human health Environmental considerationsRegulations Device options ConfigurationParameter Speaker volume ECG out dataPatient treatment protocol options During a Smart NSA pause, Pause HeartStart NSA pause type SettingsStandard NSA Monitoring duringCPR4 Child pads cartridge is installed InstalledInfants and children when an infant Yes, NoHeartstart M5066A Testing Testing and troubleshootingTroubleshooting Green Ready light is solid HeartStart tells you Troubleshooting while the HeartStart is in useTo insert a pads cartridge HeartStart tells you To stop all motion Contact Philips for service if needed Troubleshooting while the HeartStart is not in useGreen Ready light is not on Behavior Emissions test Electromagnetic conformityElectromagnetic emissions CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901