Philips M5066A owner manual Defibrillation Waveform, Category Waveform parameters, Delivered

Page 45

Philips Medical Systems

category

waveform parameters

E-3

Defibrillation Waveform

specifications

Biphasic truncated exponential. Waveform parameters are automatically adjusted as a function of patient defibrillation impedance. In the diagram at left, D is the duration of phase 1 and E is the duration of phase 2 of the waveform, F is the interphase delay (500 µs), and Ip is the peak current.

The HeartStart delivers shocks to load impedances from 25 to 180 ohms. The duration of each phase of the waveform is dynamically adjusted based on delivered charge, in order to compensate for patient impedance variations, as shown below:

 

 

adult defibrillation

 

 

 

load

phase 1

phase 2

peak

delivered

 

resistance (Ω)

duration (ms)

duration (ms)

current (A)

energy (J)

 

25

2.8

2.8

65

128

 

50

4.5

4.5

40

150

 

75

6.25

5.0

30

155

 

100

8.0

5.3

24

157

 

125

9.65

6.4

21

159

 

150

11.5

7.7

18

160

 

175

12.0

8.0

16

158

 

 

pediatric defibrillation

 

 

E

(using M5072A infant/child reduced-energy defibrillator pads)

 

load

phase 1

phase 2

peak

delivered

 

resistance (Ω)

duration (ms)

duration (ms)

current (A)

energy (J)

 

25

4.1

2.8

28

35

 

50

5.1

3.4

20

46

 

75

6.2

4.1

15

52

 

100

7.2

4.8

12

54

 

125

8.3

5.5

10

56

 

150

9.0

6.0

9

57

 

175

9.0

6.0

8

55

 

Image 45
Contents Edition M5066AIntentionally blank Pads Cartridge Handle HeartStart Defibrillator M5066AOn/Off Button. Press this Information Button. ThisIntentionally blank Quick Reference Intentionally blank Important Note HeartStartIntentionally blank Edition history About this editionCopyright Authorized EU representativeIntentionally blank Contents Appendices Description Introduction to the HeartStartSudden Cardiac Arrest Indications for UseTraining and practice National and local requirementsFor more information Package contents Setting up the HeartStartHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Overview Using the HeartStartPull the green handle Place the pads If a shock is needed Press the Shock buttonRemove the Infant/Child Pads Cartridge from its package Treating infants and childrenDo not Delay Treatment When emergency medical services arrivePhilips Medical Systems Heartstart M5066A After using the HeartStart After each useHeartStart data storage Heartstart M5066A Routine Maintenance Maintaining the HeartStartPeriodic checks Cleaning the HeartStart Troubleshooting tipsDisposing of the HeartStart REF 68-PCHAT Accessories for the HeartStartHeartstart M5066A AED Glossary of termsCPR ECGNSA Smart NSA pause SCA Glossary of symbols/controls Symbol DescriptionDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Contact with the patient Precautions Regularly for damage, according to directions Metal objects such as a bed frame or stretcherTechnical information HeartStart Defibrillator specificationsPhysical EnvironmentalCategory Sealing Controls and indicatorsCategory Waveform parameters Defibrillation WaveformDelivered Current a Energy JDisarm AED mode Shock-to-shock cycle timeNewborn 14 J/kg ECG analysis system Category Function Shockable rhythms Non-shockable rhythmsAnalysis protocol ECG analysis performance Battery M5070A Accessories specificationsCategory Specifications Battery type Prior to insertion Standby lifeEnvironmental considerations Whole and may also endanger human healthRegulations Parameter Speaker volume ConfigurationDevice options ECG out dataPatient treatment protocol options Standard NSA NSA pause type SettingsDuring a Smart NSA pause, Pause HeartStart Monitoring duringCPR4 Infants and children when an infant InstalledChild pads cartridge is installed Yes, NoHeartstart M5066A Testing and troubleshooting TestingTroubleshooting Troubleshooting while the HeartStart is in use Green Ready light is solid HeartStart tells youTo insert a pads cartridge HeartStart tells you To stop all motion Troubleshooting while the HeartStart is not in use Contact Philips for service if neededGreen Ready light is not on Behavior Electromagnetic emissions Electromagnetic conformityEmissions test CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Environmental considerations Important warnings and remindersIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901