Philips M5066A Troubleshooting while the HeartStart is in use, To insert a pads cartridge

Page 58

G-2

no spare battery is available, remove the installed battery for five seconds, then

reinsert it and run a battery insertion self-test.

If the problem continues, do not use the HeartStart. Attend to the patient,

providing CPR if needed, until Emergency Medical Services Personnel arrive.

Troubleshooting while the HeartStart is in use

(green Ready light is solid)

HeartStart tells you:

... to replace the battery immediately

... there is no cartridge installed, and

... to insert a pads cartridge

... to press the pads firmly to the skin

... to make sure the pads have been removed from the liner

... the pads should not be touching the patient’s clothing.

... to insert new pads cartridge

possible cause

The battery is nearly depleted. The defibrillator will turn off if a new battery is not inserted.

The pads cartridge has been removed.

The pads cartridge has been damaged.

The pads are not properly applied to the patient.

The pads are not making good contact with the patient's bare chest because of moisture or excessive hair.

The pads are touching each other.

The pads may not have been removed from the liner or may be on the patient’s clothing.

The pads cartridge has been opened and the pads peeled off the liner, but the pads have not been successfully attached to the patient. There may be a problem with the pads cartridge.

recommended action

Replace the battery with a new battery immediately.

Insert a new pads cartridge.

Make sure that the pads are sticking completely to the patient’s skin.

If the pads are not sticking, dry the patient's chest and shave or clip any excessive chest hair.

Reposition the pads.

Make sure the pads are not on the liner or the patient’s clothing.

If the voice instruction continues after you do these things, insert another pads cartridge.

Replace the damaged pads cartridge. Pull up the handle on the cartridge cover, and replace pads on patient with new pads to continue with the rescue.

Philips Medical Systems

HEARTSTART M5066A

Image 58
Contents M5066A EditionIntentionally blank On/Off Button. Press this HeartStart Defibrillator M5066APads Cartridge Handle Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank Copyright About this editionEdition history Authorized EU representativeIntentionally blank Contents Appendices Sudden Cardiac Arrest Introduction to the HeartStartDescription Indications for UseNational and local requirements Training and practiceFor more information Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A After each use After using the HeartStartHeartStart data storage Heartstart M5066A Maintaining the HeartStart Routine MaintenancePeriodic checks Troubleshooting tips Cleaning the HeartStartDisposing of the HeartStart Accessories for the HeartStart REF 68-PCHATHeartstart M5066A CPR Glossary of termsAED ECGNSA Smart NSA pause SCA Symbol Description Glossary of symbols/controlsDefibrillator and opens pads cartridge for use Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Regularly for damage, according to directions PrecautionsContact with the patient Metal objects such as a bed frame or stretcherPhysical HeartStart Defibrillator specificationsTechnical information EnvironmentalControls and indicators Category SealingDelivered Defibrillation WaveformCategory Waveform parameters Current a Energy JShock-to-shock cycle time Disarm AED modeNewborn 14 J/kg Category Function Shockable rhythms Non-shockable rhythms ECG analysis systemAnalysis protocol ECG analysis performance Category Specifications Battery type Accessories specificationsBattery M5070A Prior to insertion Standby lifeWhole and may also endanger human health Environmental considerationsRegulations Device options ConfigurationParameter Speaker volume ECG out dataPatient treatment protocol options During a Smart NSA pause, Pause HeartStart NSA pause type SettingsStandard NSA Monitoring duringCPR4 Child pads cartridge is installed Installed Infants and children when an infant Yes, NoHeartstart M5066A Testing Testing and troubleshootingTroubleshooting Green Ready light is solid HeartStart tells you Troubleshooting while the HeartStart is in useTo insert a pads cartridge HeartStart tells you To stop all motion Contact Philips for service if needed Troubleshooting while the HeartStart is not in useGreen Ready light is not on Behavior Emissions test Electromagnetic conformityElectromagnetic emissions CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901