Philips M5066A owner manual Smart NSA pause

Page 35

Philips Medical Systems

 

B-3

 

 

On/Off button stops the battery insertion self-test that automatically runs when a

 

 

battery is inserted.

 

pads

See “SMART pads.”

 

patient care pause

A defined pause to allow patient assessment, treatment, and/or CPR. See

B

 

“NSA pause” and “protocol pause.”

 

periodic self-tests

Daily, weekly, and monthly tests automatically conducted by the HeartStart

 

 

Defibrillator when it is in its standby mode. The tests monitor many key functions

 

 

and parameters of the defibrillator, including battery capacity, pads cartridge

 

 

readiness, and the state of its internal circuitry.

 

protocol

A sequence of operations performed by the HeartStart Defibrillator to direct

 

 

patient care in the AED mode.

 

protocol pause

A pause provided by the HeartStart Defibrillator after a shock series, during which

 

 

the responder can administer CPR. The defibrillator does not conduct background

 

 

monitoring of the patient’s heart rhythm during this pause.

 

Ready light

A green LED showing the readiness for use of the HeartStart Defibrillator. A

 

 

blinking Ready light means the defibrillator is ready for use; a solid Ready light

 

 

means the defibrillator is being used.

 

rhythm analysis

See “SMART analysis.”

 

Shock button

An orange button with a lightning bolt symbol on it, located on the front of the

 

 

HeartStart Defibrillator. The Shock button flashes when a shock is advised. You

 

 

must press the button for the shock to be delivered.

 

shockable rhythm

A heart rhythm that the HeartStart Defibrillator determines is appropriate for

 

 

defibrillation, such as ventricular fibrillation and some ventricular tachycardias

 

 

associated with sudden cardiac arrest.

 

shock series interval

A configurable interval between shocks, used by the HeartStart Defibrillator to

 

 

decide if the shocks are part of the same shock series.

 

SMART analysis

The proprietary algorithm used by the HeartStart Defibrillator to analyze the

 

 

patient’s heart rhythm and determine whether the rhythm is shockable.

 

SMART biphasic

The patented, low-energy defibrillation shock waveform used by the HeartStart

 

waveform

Defibrillator. It is an impedance-compensated biphasic waveform. Used with the

 

 

Adult SMART Pads, it delivers 150 Joules, nominal, into a 50 ohm load; used with

 

 

the Infant/Child SMART Pads, it delivers 50 Joules, nominal, into a 50 ohm load.

 

SMART NSA pause

See “NSA pause.”

 

SMART Pads

The adhesive pads, supplied in a cartridge, used with the HeartStart Defibrillator.

 

 

Pulling the handle on the cartridge turns on the defibrillator and opens the

 

Image 35
Contents Edition M5066AIntentionally blank Information Button. This HeartStart Defibrillator M5066APads Cartridge Handle On/Off Button. Press thisIntentionally blank Quick Reference Intentionally blank Important Note HeartStartIntentionally blank Authorized EU representative About this editionEdition history CopyrightIntentionally blank Contents Appendices Indications for Use Introduction to the HeartStartDescription Sudden Cardiac ArrestFor more information Training and practiceNational and local requirements Package contents Setting up the HeartStartHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Overview Using the HeartStartPull the green handle Place the pads If a shock is needed Press the Shock buttonRemove the Infant/Child Pads Cartridge from its package Treating infants and childrenDo not Delay Treatment When emergency medical services arrivePhilips Medical Systems Heartstart M5066A HeartStart data storage After using the HeartStartAfter each use Heartstart M5066A Periodic checks Routine MaintenanceMaintaining the HeartStart Disposing of the HeartStart Cleaning the HeartStartTroubleshooting tips REF 68-PCHAT Accessories for the HeartStartHeartstart M5066A ECG Glossary of termsAED CPRNSA Smart NSA pause SCA Defibrillator and opens pads cartridge for use Glossary of symbols/controlsSymbol Description Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Metal objects such as a bed frame or stretcher PrecautionsContact with the patient Regularly for damage, according to directionsEnvironmental HeartStart Defibrillator specificationsTechnical information PhysicalCategory Sealing Controls and indicatorsCurrent a Energy J Defibrillation WaveformCategory Waveform parameters DeliveredNewborn 14 J/kg Disarm AED modeShock-to-shock cycle time Analysis protocol ECG analysis systemCategory Function Shockable rhythms Non-shockable rhythms ECG analysis performance Prior to insertion Standby life Accessories specificationsBattery M5070A Category Specifications Battery typeRegulations Environmental considerationsWhole and may also endanger human health ECG out data ConfigurationParameter Speaker volume Device optionsPatient treatment protocol options Monitoring during NSA pause type SettingsStandard NSA During a Smart NSA pause, Pause HeartStartCPR4 Yes, No InstalledInfants and children when an infant Child pads cartridge is installedHeartstart M5066A Troubleshooting Testing and troubleshootingTesting To insert a pads cartridge Troubleshooting while the HeartStart is in useGreen Ready light is solid HeartStart tells you HeartStart tells you To stop all motion Green Ready light is not on Behavior Troubleshooting while the HeartStart is not in useContact Philips for service if needed Cispr Electromagnetic conformityElectromagnetic emissions Emissions testElectromagnetic immunity Separation distance according to frequency of transmitter m Environmental considerations Important warnings and remindersIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901
Related manuals
Manual 68 pages 2.08 Kb

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.