Philips M5066A owner manual Electromagnetic immunity

Page 62

H-2

Electromagnetic immunity

immunity test

electrostatic discharge (ESD)

IEC 61000-4-2

power frequency (50/60 Hz) magnetic field IEC 61000-4-8

radiated RF IEC 61000-4-3

IEC 60601

compliance level

electromagnetic environment -

test level

guidance

 

± 6 kV contact

± 6 kV contact

There are no special requirements

± 8 kV air

± 8 kV air

with respect to electrostatic

 

 

discharge.a

3 A/m

3 A/m

Power frequency magnetic fields

 

 

should be at levels characteristic of a

 

 

typical location in a typical

 

 

commercial/hospital environment.

 

 

There are no special requirements for

 

 

non-commercial/non-hospital

 

 

environments.

10 V/m

20 V/m

Portable and mobile RF

80 MHz to 2.5

 

communications equipment should be

GHz

 

used no closer to any part of the

 

 

HeartStart, including cables, than is

 

 

absolutely necessary.b,c The

 

 

recommended separation distances

 

 

for various transmitters and the AED

 

 

are shown in the following table.

 

 

Interference may occur in the

 

 

vicinity of equipment marked

 

 

with the following symbol:

Philips Medical Systems

NOTE 1. At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a.Generally, AEDs are sometimes susceptible to interference generated by patient and/or responder motion in environments in which a high static electric field is present (e.g., low humidity, synthetic carpets, etc.). As a safety measure, Philips AEDs incorporate a patented method to sense possible corruption of the ECG signal by such interference and to respond by directing the user to stop all motion. In these cases, it is important to minimize movement in the vicinity of the patient during rhythm analysis in order to ensure that the signal being analyzed accurately reflects the patient’s underlying heart rhythm.

b.The ISM (industrial, scientific and medial) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz.

c.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the HeartStart is used exceeds the applicable RF compliance level above, the HeartStart should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the HeartStart.

HEARTSTART M5066A

Image 62
Contents M5066A EditionIntentionally blank On/Off Button. Press this HeartStart Defibrillator M5066APads Cartridge Handle Information Button. ThisIntentionally blank Quick Reference Intentionally blank HeartStart Important NoteIntentionally blank Copyright About this editionEdition history Authorized EU representativeIntentionally blank Contents Appendices Sudden Cardiac Arrest Introduction to the HeartStartDescription Indications for UseFor more information Training and practiceNational and local requirements Setting up the HeartStart Package contentsHeartstart M5066A Pads Cartridge, place it in the storage area in the case Recommended accessories Using the HeartStart OverviewPull the green handle Place the pads Press the Shock button If a shock is neededTreating infants and children Remove the Infant/Child Pads Cartridge from its packageWhen emergency medical services arrive Do not Delay TreatmentPhilips Medical Systems Heartstart M5066A HeartStart data storage After using the HeartStartAfter each use Heartstart M5066A Periodic checks Routine MaintenanceMaintaining the HeartStart Disposing of the HeartStart Cleaning the HeartStartTroubleshooting tips Accessories for the HeartStart REF 68-PCHATHeartstart M5066A CPR Glossary of termsAED ECGNSA Smart NSA pause SCA Defibrillator and opens pads cartridge for use Glossary of symbols/controlsSymbol Description Lithium manganese dioxide battery Cartridge contents one set of two defibrillation pads Symbol Description Patient handling Regularly for damage, according to directions PrecautionsContact with the patient Metal objects such as a bed frame or stretcherPhysical HeartStart Defibrillator specificationsTechnical information EnvironmentalControls and indicators Category SealingDelivered Defibrillation WaveformCategory Waveform parameters Current a Energy JNewborn 14 J/kg Disarm AED modeShock-to-shock cycle time Analysis protocol ECG analysis systemCategory Function Shockable rhythms Non-shockable rhythms ECG analysis performance Category Specifications Battery type Accessories specificationsBattery M5070A Prior to insertion Standby lifeRegulations Environmental considerationsWhole and may also endanger human health Device options ConfigurationParameter Speaker volume ECG out dataPatient treatment protocol options During a Smart NSA pause, Pause HeartStart NSA pause type SettingsStandard NSA Monitoring duringCPR4 Child pads cartridge is installed InstalledInfants and children when an infant Yes, NoHeartstart M5066A Troubleshooting Testing and troubleshootingTesting To insert a pads cartridge Troubleshooting while the HeartStart is in useGreen Ready light is solid HeartStart tells you HeartStart tells you To stop all motion Green Ready light is not on Behavior Troubleshooting while the HeartStart is not in useContact Philips for service if needed Emissions test Electromagnetic conformityElectromagnetic emissions CisprElectromagnetic immunity Separation distance according to frequency of transmitter m Important warnings and reminders Environmental considerationsIntentionally blank Intentionally blank Intentionally blank Intentionally blank Intentionally blank REF M5066-91901
Related manuals
Manual 68 pages 2.08 Kb

M5066A specifications

The Philips M5066A is a state-of-the-art multi-parameter patient monitor designed to enhance healthcare delivery in various clinical settings. This versatile monitoring solution is widely utilized in hospitals, especially in critical care units and during transport, owing to its robust features and reliable performance.

One of the main features of the M5066A is its ability to monitor multiple vital signs simultaneously, including ECG, heart rate, invasive and non-invasive blood pressure, respiration rate, temperature, and saturation levels. This multi-parameter capability allows healthcare professionals to gain a comprehensive view of a patient’s health status, enabling timely interventions and improved patient outcomes.

The M5066A incorporates advanced technologies such as Philips’ proprietary IntelliVue technology, which enhances the accuracy and reliability of vital sign measurements. The monitor uses sophisticated algorithms to filter out noise and interferences, ensuring that the data presented to caregivers is both precise and actionable. The device also supports a range of connectivity options, allowing for seamless integration with hospital information systems and electronic medical records (EMRs). This connectivity facilitates the easy transfer and management of patient data, promoting better coordination of care.

Another significant characteristic of the M5066A is its user-friendly interface. The large, high-resolution display provides clear visuals of vital sign trends and parameters, enabling quick assessments by healthcare professionals. Customizable alarms and alerts can be programmed to suit specific patient needs, ensuring that caregivers are notified of critical changes in a timely manner.

The M5066A is also designed with patient comfort and safety in mind. The device's compact and lightweight design facilitates easy transport, while its durable construction ensures reliability in demanding environments. Furthermore, the monitor's battery backup allows for continuous monitoring even during power outages or when moving patients between departments.

In conclusion, the Philips M5066A multi-parameter patient monitor is a sophisticated tool that enhances patient monitoring by providing accurate, real-time data. Its advanced technologies, comprehensive monitoring capabilities, user-friendly interface, and focus on safety and comfort make it an invaluable asset in any clinical setting, helping healthcare providers deliver the highest standard of care.