The pH ARC Sensor register structure uses the following physical units used for Primary or Secondary Measurement Channels.

Bit #

Hex code

Physical unit

Start register. (8 ASCII characters, length 4 registers,

 

 

 

Type 3, read for U/A/S)

0 (LSB)

0x00000001

none

1920

1

0x00000002

K

1924

2

0x00000004

°C

1928

3

0x00000008

°F

1932

4

0x00000010

%-vol

1936

5

0x00000020

%-sat

1940

6

0x00000040

ug/l ppb

1944

7

0x00000080

mg/l ppm

1948

8

0x00000100

g/l

1952

9

0x00000200

uS/cm

1956

10

0x00000400

mS/cm

1960

11

0x00000800

1/cm

1964

12

0x00001000

pH

1968

13

0x00002000

mV/pH

1972

14

0x00004000

kOhm

1976

15

0x00008000

MOhm

1980

16

0x00010000

pA

1984

17

0x00020000

nA

1988

18

0x00040000

uA

1992

19

0x00080000

mA

1996

20

0x00100000

uV

2000

21

0x00200000

mV

2004

22

0x00400000

V

2008

23

0x00800000

mbar

2012

24

0x01000000

Pa

2016

25

0x02000000

Ohm

2020

26

0x04000000

%/°C

2024

27

0x08000000

°

2028

28

0x10000000

not used

2032

29

0x20000000

not used

2036

30

0x40000000

not used

2040

31 (MSB)

0x80000000

SPECIAL

2044

Figure 2.5.1.4: Definition of physical units used for PMC and SMC.

Command: Unit text

Modbus address: 1968

Length: 4

Type: 3

Read

Parameter:

Text

 

 

 

 

Format:

Character

 

 

 

 

Value:

pH

 

 

 

 

Figure 2.5.1.5: Example to read the physical unit in plain text ASCII in register 1968

624300/01

page 19 / 67

Image 19
Contents PH ARC Sensors Important Notice Sensor Identification and Information Configuration of the Serial RS485 InterfaceConfiguration of the Analog Interfaces Configuration of the MeasurementVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 IntroductionEPHUM011 0x03 Operator levels and PasswordsReading / Setting Operator Level General0x0B 18111978 Changing Passwords for Operator Level0x30 0x0C 18111978Reading the Device Address Limits Configuration of the serial RS485 InterfaceDevice Address Reading and Writing the Device AddressReading and Writing the Baud Rate Baud RateReading the Baud Rate Limits 0x07 Configuration of the Analog InterfacesAvailable Analog Interfaces Available Analog Interface ModesMA interface #2 Selection of an Analog Interface ModeDescription of the Analog Interfaces 1 MA interface #10x21 Configuration of the 4-20 mA Interface0x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 0x001000 Reading the Selected Physical Unit for Analog InterfaceDefining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial 0x201000 Primary Measurement Channel 1 pHDescription of PMC1 Selecting the Physical Unit for PMC10x1000 02503 0x00 Reading the measurement value of PMC10x200000 175.9922 0x00 954.6541 0x06 Primary Measurement Channel 6 TemperatureDescription of PMC6 Selecting the Physical Unit for PMC60x04 24.35834 0x00 130 Input of an Externally Measured TemperatureDefinition of the Measurement Status for PMC1 / PMC6 Reading the measurement value of PMC6Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 Description of PA12 Moving Average R 3 PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Available Calibration Points CalibrationCP2 Definitions of Calibration PointsCalibration Points 1 and 2 Standard Calibration CP1CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration Status0x00000180 0x00001000 0x00000080 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x14000000 0x00001000 0x01000000 0x000010000x08000000 0x00001000 0x0A000000 0x000010000x10000000 0x00001000 0x00000004 29.93368 379.5167 Currently active Calibration Parameters part0x00000004 24.35184 23.78 0x00000004 24.37691 16.45607782 59.47631 298.15 003401 179.927 297.1378006804 099747 296.6901 10469 298.3302Special Commands for Calibration with Visical 0x0000003F Calibration StandardsAvailable Sets of Calibration Standards LSB 0x00000001 Set180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersHistory of Warnings Currently Active WarningsDefinition of Warnings Currently Active Errors ErrorsHistory of Errors Definition of Errors SIP Reading Definition of SIP and CIPCIP 120 130 Reading the Sensor’s Quality IndicatorEPHFI010 Sensor Identification and InformationGeneral Information Sensor Identification0001001 Free User Memory SpaceFreeuserspace 732255 System CommandsRecall Sensor’s Factory Settings Hello WorldPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011