Hamilton Electronics EPHUM011 manual Definition of the Measurement Status for PMC1 / PMC6

Page 23

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.5.3.3Reading the measurement value of PMC6

Register 2410 is also used to read the measurement values of PMC6.

Start

Num-

Reg1 /

Reg3 /

Reg5 /

Reg7 /

Reg9 /

Modbus

Read

Write

reg.

ber of

Reg2

Reg4

Reg6

Reg8

Reg10

function

access

access

 

reg.

 

 

 

 

 

code

 

 

2410

10

Selected

Measure

Measure

Min

Max

3, 4

U/A/S

none

 

 

physical

-ment

-ment

allowed

allowed

 

 

 

 

 

unit

value of

status (2)

value (1)

value (1)

 

 

 

 

 

 

PMC6 (1)

 

 

 

 

 

 

Figure 2.5.3.3.1: Definition of register 2410. Measurement value of PMC6.

(1)Value is always in the physical unit defined in register 2410.

(2)For definition of the status see chapter 2.5.4. All bits set to zero means: no problem.

Command: PMC6 read

 

Modbus address: 2410

Length: 10

Type: 3

Read

Parameter:

Unit

Value

 

Status

Min limit

 

Max limit

 

Format:

Hex

Float

 

Hex

Float

 

Float

 

Value:

0x04

24.35834

 

0x00

-20

 

130

 

Figure 2.5.3.3.2: Example to read register 2410. Physical unit is set to °C (0x04), PMC6 is 24.35834 °C, Status is 0x00, Min allowed value is -20 °C, Max allowed value is 130 °C.

For definition of the measurement status see chapter 2.5.4.

Attention:

You cannot read selectively the registers 3 and 4 for the measurement value only. You have to read the entire length of the command (10 registers) and extract the desired information.

2.5.3.4Input of an Externally Measured Temperature

Unlike to the VISIFERM DO, this feature is not available for pH ARC Sensors.

2.5.4 Definition of the Measurement Status for PMC1 / PMC6

This is the definition of the status registers read in registers 2090 (PMC1) and 2410 (PMC6):

Bit #

Hex code

Description

0 (LSB)

0x01

Temperature out of measurement range (see chapter 2.8.1)

1

0x02

Temperature out of operating range (see chapter 2.8.1)

2

0x04

Calibration status not zero (see chapter 2.7.4)

3

0x08

Warning not zero (see chapter 2.8.3)

4

0x10

Error not zero (see chapter 2.8.4)

Figure 2.5.4.1: Definition of measurement status for Primary Measurement Channels.

624300/01

page 23 / 67

Image 23
Contents PH ARC Sensors Important Notice Sensor Identification and Information Configuration of the Serial RS485 InterfaceConfiguration of the Analog Interfaces Configuration of the MeasurementEPHUM011 IntroductionVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 0x03 Operator levels and PasswordsReading / Setting Operator Level General0x0B 18111978 Changing Passwords for Operator Level0x30 0x0C 18111978Reading the Device Address Limits Configuration of the serial RS485 InterfaceDevice Address Reading and Writing the Device AddressReading the Baud Rate Limits Baud RateReading and Writing the Baud Rate 0x07 Configuration of the Analog InterfacesAvailable Analog Interfaces Available Analog Interface ModesMA interface #2 Selection of an Analog Interface ModeDescription of the Analog Interfaces 1 MA interface #10x01 Configuration of the 4-20 mA Interface0x21 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Defining the Measurement Values for 4, 12 and 20 mA Output Reading the Selected Physical Unit for Analog Interface0x001000 PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial 0x201000 Primary Measurement Channel 1 pHDescription of PMC1 Selecting the Physical Unit for PMC10x200000 175.9922 0x00 954.6541 Reading the measurement value of PMC10x1000 02503 0x00 0x06 Primary Measurement Channel 6 TemperatureDescription of PMC6 Selecting the Physical Unit for PMC60x04 24.35834 0x00 130 Input of an Externally Measured TemperatureDefinition of the Measurement Status for PMC1 / PMC6 Reading the measurement value of PMC6Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 Selecting the Physical Unit and Writing the Value for PA12 3 PA12 Moving Average RDescription of PA12 Moving Average R Reading all Values for PA12 Available Calibration Points CalibrationCP2 Definitions of Calibration PointsCalibration Points 1 and 2 Standard Calibration CP1CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration StatusReading the Calibration Status of CP6 Product Calibration 0x00000080 0x000010000x00000180 0x00001000 0x14000000 0x00001000 0x01000000 0x000010000x08000000 0x00001000 0x0A000000 0x000010000x10000000 0x00001000 0x00000004 29.93368 379.5167 Currently active Calibration Parameters part0x00000004 24.35184 23.78 0x00000004 24.37691 16.45607782 59.47631 298.15 003401 179.927 297.1378006804 099747 296.6901 10469 298.3302Special Commands for Calibration with Visical 0x0000003F Calibration StandardsAvailable Sets of Calibration Standards LSB 0x00000001 Set180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersDefinition of Warnings Currently Active WarningsHistory of Warnings History of Errors ErrorsCurrently Active Errors Definition of Errors CIP Reading Definition of SIP and CIPSIP 120 130 Reading the Sensor’s Quality IndicatorEPHFI010 Sensor Identification and InformationGeneral Information Sensor IdentificationFreeuserspace Free User Memory Space0001001 732255 System CommandsRecall Sensor’s Factory Settings Hello WorldPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011