Hamilton Electronics EPHUM011 003401 179.927 297.1378, 006804 099747 296.6901, 10469 298.3302

Page 49

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.7 Currently active Calibration Parameters part 3

In register 5520, 5528 and 5560 the pH value of the used calibration standards, the electrical potential of the pH sensor and the temperature upon calibration are stored.

Start

Number

Reg1 /

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

of

Reg2

Reg4

Reg6

Reg8

function

access

access

 

registers

 

 

 

 

code

 

 

5520

8

pH value of

Electrical

Electrical

free

3, 4

A/S

none

 

 

calibration

potential at

potential at

 

 

 

 

 

 

standard at

CP1

CP1

 

 

 

 

 

 

CP1 [pH]

[mV]

[mV]

 

 

 

 

5528

8

pH value of

Electrical

Electrical

free

3, 4

A/S

none

 

 

calibration

potential at

potential at

 

 

 

 

 

 

standard at

CP2

CP2

 

 

 

 

 

 

CP2 [pH]

[mV]

[mV]

 

 

 

 

5560

8

pH value of

Electrical

Electrical

free

3, 4

A/S

none

 

 

product at

potential at

potential at

 

 

 

 

 

 

CP6 [pH]

CP6 [mV]

CP6 [mV]

 

 

 

 

Figure 2.7.7.1: Definition of register 5520, 5528 and 5560.

Command: Act calibration CP1

Modbus address: 5520

 

Length: 8

Type: 3

Read

Parameter:

pH CP1 [pH]

Voltage CP1 [mV]

Temp

CP1 [K]

free

 

Format:

Float

Float

Float

 

Float

 

Value:

4.003401

179.927

297.1378

0

 

Figure 2.7.7.2:

Example to read the

actual calibration values

of CP1.

 

 

 

 

 

 

 

 

 

Command: Act calibration CP2

Modbus address: 5528

 

Length: 8

Type: 3

Read

Parameter:

pH CP2 [pH]

Voltage CP2 [mV]

Temp

CP2 [K]

free

 

Format:

Float

Float

Float

 

Float

 

Value:

7.006804

3.099747

296.6901

0

 

Figure 2.7.7.3: Example to read the actual calibration values of CP2.

Command: Act calibration CP6

Modbus address: 5560

Length: 8

Type: 3

Read

Parameter:

pH CP6 [pH]

Voltage CP6 [mV]

Temp

CP6 [K]

free

 

Format:

Float

Float

Float

 

Float

 

Value:

7.1

5.10469

298.3302

0

 

Figure 2.7.7.4: Example to read the actual calibration values of CP6.

2.7.8 Currently active Calibration Parameters part 4

For standard calibration (CP1 / CP2) register 5448 documents offset at pH7 and slope:

Start

Number

Reg1 / Reg2

 

Reg3 / Reg4

Reg5 / Reg6

 

Modbus

Read

 

Write

register

of

(Float)

 

(Float)

(Float)

 

 

function

access

 

access

 

registers

 

 

 

 

 

 

 

code

 

 

 

 

5448

6

 

Offset at

 

Slope (25 °C)

Reference

 

3, 4

 

U/A/S

 

none

 

 

 

pH 7 [mV]

 

[mV/pH]

temperature [K]

 

 

 

 

 

 

 

Figure 2.7.8.1: Definition of register 5448.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Command: Calculated cali values

 

Modbus address: 5448

Length: 6

 

Type: 3

Read

 

Parameter:

Offset at pH 7

Slope

 

Ref temp

 

 

 

 

 

 

 

 

[mV]

 

[mV/pH]

 

[K]

 

 

 

 

 

 

 

 

Format:

 

Float

 

Float

 

Float

 

 

 

 

 

 

 

 

Value:

 

3.607782

-59.47631

 

298.15

 

 

 

 

 

 

 

 

Figure 2.7.8.2: Example to read register 5448: offset at pH 7 is 3.6 mV; slope is -59.5 mV/pH; reference temperature is 298.15 K (=25°C)

624300/01

page 49 / 67

Image 49
Contents PH ARC Sensors Important Notice Configuration of the Analog Interfaces Configuration of the Serial RS485 InterfaceConfiguration of the Measurement Sensor Identification and InformationVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 IntroductionEPHUM011 Reading / Setting Operator Level Operator levels and PasswordsGeneral 0x030x30 Changing Passwords for Operator Level0x0C 18111978 0x0B 18111978Device Address Configuration of the serial RS485 InterfaceReading and Writing the Device Address Reading the Device Address LimitsReading and Writing the Baud Rate Baud RateReading the Baud Rate Limits Available Analog Interfaces Configuration of the Analog InterfacesAvailable Analog Interface Modes 0x07Description of the Analog Interfaces 1 Selection of an Analog Interface ModeMA interface #1 MA interface #20x21 Configuration of the 4-20 mA Interface0x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 0x001000 Reading the Selected Physical Unit for Analog InterfaceDefining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial Description of PMC1 Primary Measurement Channel 1 pHSelecting the Physical Unit for PMC1 0x2010000x1000 02503 0x00 Reading the measurement value of PMC10x200000 175.9922 0x00 954.6541 Description of PMC6 Primary Measurement Channel 6 TemperatureSelecting the Physical Unit for PMC6 0x06Definition of the Measurement Status for PMC1 / PMC6 Input of an Externally Measured TemperatureReading the measurement value of PMC6 0x04 24.35834 0x00 130Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 Description of PA12 Moving Average R 3 PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Available Calibration Points CalibrationCalibration Points 1 and 2 Standard Calibration Definitions of Calibration PointsCP1 CP2CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration Status0x00000180 0x00001000 0x00000080 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x08000000 0x00001000 0x01000000 0x000010000x0A000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 0x00000004 24.35184 23.78 Currently active Calibration Parameters part0x00000004 24.37691 16.45 0x00000004 29.93368 379.5167006804 099747 296.6901 003401 179.927 297.137810469 298.3302 607782 59.47631 298.15Special Commands for Calibration with Visical Available Sets of Calibration Standards Calibration StandardsLSB 0x00000001 Set 0x0000003F180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersHistory of Warnings Currently Active WarningsDefinition of Warnings Currently Active Errors ErrorsHistory of Errors Definition of Errors SIP Reading Definition of SIP and CIPCIP 120 130 Reading the Sensor’s Quality IndicatorGeneral Information Sensor Identification and InformationSensor Identification EPHFI0100001001 Free User Memory SpaceFreeuserspace Recall Sensor’s Factory Settings System CommandsHello World 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011