Hamilton Electronics EPHUM011 manual 3 PA12 Moving Average R, Description of PA12 Moving Average R

Page 30

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.6.3 PA12: Moving Average R

pH ARC Sensors allow to have a separate moving average on secondary measurement values:

Glass resistance

Reference resistance

The moving average can be applied on 1 to 16 3-s measurement values. The default value is 4. Especially if high resistances are measured, it is recommended to choose a higher moving average.

2.6.3.1Description of PA12 (Moving Average R)

In register 3456, a plain text ASCII description of PA12 is given.

Start

Number of

Reg1 … Reg8

Modbus

Read

Write

register

registers

16 ASCII characters

function code

access

access

3456

8

Description of PA12

3, 4

U/A/S

none

Figure 2.6.3.1.1: Definition of register 3456.

Command: Moving average text

Modbus address: 3456

Length: 8

Type: 3

Read

Parameter:

Text

 

 

 

 

 

 

Format:

Character

 

 

 

 

 

 

Value:

Moving average R

 

 

 

 

 

 

Figure 2.6.3.1.2: Example to read the description for “Moving average R“.

2.6.3.2Selecting the Physical Unit and Writing the Value for PA12

In register 3464, the available physical units for PA12 are defined.

Start

Number of

Reg1 / Reg2

Modbus

 

Read

Write

register

registers

(bitwise defined)

function code

 

access

access

3464

2

 

Available physical units for PA12

3, 4

 

U/A/S

none

Figure 2.6.3.2.1: Definition of register 3368.

 

 

 

 

 

 

 

 

 

 

 

 

Command: Moving average av. units

Modbus address: 3464

Length: 2

Type: 3

Read

Parameter:

Units

 

 

 

 

 

 

 

 

Format:

 

Hex

 

 

 

 

 

 

 

 

Value:

 

0x01

 

 

 

 

 

 

 

 

Figure 2.6.3.2.2: Example to read the available physical units for PA12. The only one available here is “none”(0x01). For the definition of the physical units see chapter 2.5.1.

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Modbus

Read

Write

register

registers

(bitwise defined)

 

function code

access

access

3466

4

Select physical unit

Value for PA12

16

none

S

 

 

for PA12

(1-16, default: 4)

 

 

 

Figure 2.6.3.2.3: Definition of register 3466. Only one bit for the physical unit can be set. PA12 can be set to the values 1-16.

624300/01

page 30 / 67

Image 30
Contents PH ARC Sensors Important Notice Configuration of the Measurement Configuration of the Serial RS485 InterfaceConfiguration of the Analog Interfaces Sensor Identification and InformationIntroduction Visiferm do Modbus RTU Programmer’s Manualp/n 624179/01EPHUM011 General Operator levels and PasswordsReading / Setting Operator Level 0x030x0C 18111978 Changing Passwords for Operator Level0x30 0x0B 18111978Reading and Writing the Device Address Configuration of the serial RS485 InterfaceDevice Address Reading the Device Address LimitsBaud Rate Reading and Writing the Baud RateReading the Baud Rate Limits Available Analog Interface Modes Configuration of the Analog InterfacesAvailable Analog Interfaces 0x07MA interface #1 Selection of an Analog Interface ModeDescription of the Analog Interfaces 1 MA interface #2Configuration of the 4-20 mA Interface 0x210x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Reading the Selected Physical Unit for Analog Interface 0x001000Defining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 Reading the Internally Measured Output Current 99186 99742Measurement Definition of Measurement Channels and Physical UnitsSpecial Selecting the Physical Unit for PMC1 Primary Measurement Channel 1 pHDescription of PMC1 0x201000Reading the measurement value of PMC1 0x1000 02503 0x000x200000 175.9922 0x00 954.6541 Selecting the Physical Unit for PMC6 Primary Measurement Channel 6 TemperatureDescription of PMC6 0x06Reading the measurement value of PMC6 Input of an Externally Measured TemperatureDefinition of the Measurement Status for PMC1 / PMC6 0x04 24.35834 0x00 130Secondary Measurement Channels Description of SMCReading the measurement value of SMC 0x8000 247.56Configuration of the Measurement Available Parameters2 PA9 Moving Average Effect of Moving Average n=1 vs n=16Description of PA9 Moving Average Selecting the Physical Unit and Writing the Value for PA9Reading all Values for PA9 3 PA12 Moving Average R Description of PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Calibration Available Calibration PointsCP1 Definitions of Calibration PointsCalibration Points 1 and 2 Standard Calibration CP2Calibration Point 6 Product Calibration CP6Calibration Procedure Calibration at CP1 and CP2 Standard CalibrationPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 100 Calibration at CP6 Product Calibration200 Definition of the commands for product calibration Product calibration Initial measurementProduct calibration Cancel Product calibration AssignmentProduct calibration Restore standard calibration Product calibration Restore product calibrationReading the Calibration Status Reading the Calibration Status of CP1 and CP20x00000080 0x00001000 0x00000180 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x0A000000 0x00001000 0x01000000 0x000010000x08000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 0x00000004 24.37691 16.45 Currently active Calibration Parameters part0x00000004 24.35184 23.78 0x00000004 29.93368 379.516710469 298.3302 003401 179.927 297.1378006804 099747 296.6901 607782 59.47631 298.15Special Commands for Calibration with Visical LSB 0x00000001 Set Calibration StandardsAvailable Sets of Calibration Standards 0x0000003FDefinitions for Individual Sets of Calibration Standards 1800x0FFF0FFF 0x02480FFF 0x05540FFFDIN Nist StandardMettler Toledo Radiometer Temperature Ranges Sensor StatusOperating Hours and Counters 168.3667Currently Active Warnings History of WarningsDefinition of Warnings Errors Currently Active ErrorsHistory of Errors Definition of Errors Reading Definition of SIP and CIP SIPCIP Reading the Sensor’s Quality Indicator 120 130Sensor Identification Sensor Identification and InformationGeneral Information EPHFI010Free User Memory Space 0001001Freeuserspace Hello World System CommandsRecall Sensor’s Factory Settings 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011