pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
What happens to the pH ARC Sensor’s calibration function upon product calibration (CP6)?
A product calibration adds an offset to the linear calibration function defined by the standard calibration at CP1 and CP2.
|
|
| Product calibration of ARC pH sensors |
| |||
| 200 |
|
|
| Standard Calibration function |
| |
|
|
|
|
|
| ||
|
|
|
|
| Product Calibration function |
| |
| 150 |
|
|
|
|
| |
[mV] | 100 |
|
|
|
|
| |
50 |
|
|
|
|
| ||
potential | CP1 |
| CP6 |
|
| ||
|
|
|
| ||||
0 |
|
|
|
|
| ||
5 | 6 | 7 | 8 | 9 | 10 | ||
Electrical | |||||||
|
|
|
|
| |||
|
|
|
|
| |||
|
|
|
|
|
| ||
|
|
|
| CP2 |
| ||
|
|
|
|
|
| ||
|
|
|
|
|
| ||
|
|
|
| pH |
|
|
Figure 2.7.3.2.1: Effect of the product calibration CP6 on an existing standard calibration function defined by CP1 and CP2.
The operator starts with a Standard Calibration with calibration points CP1 and CP2:
CP1: pH value of calibration standard: 6 | electrical potential: 64.2 mV |
CP2: pH value of calibration standard: 9.21 | electrical potential: |
The sensor internally calculates the calibration function, using the calibration points CP1 and CP2. The resulting calibration function, compensated to the standard temperature 25°C, is shown as a straight line. The calibration function is described by two parameters: the offset at pH 7 and the slope.
Some weeks later, the operator believes that the Standard Calibration function is not correct anymore. As the process is running and he is not able to perform a standard calibration under defined conditions in the lab, he decides to perform a product calibration CP6, in other words adjusting the standard calibration function to the process conditions:
CP6: pH value of product: 7.2 electrial potential: 5 mV
The sensor internally adds an offset to the calibration curve. The slope remains unchanged.
Another special feature of this calibration point is to switch off and back on again a product calibration. These functions are called “restore standard calibration”and “restore product calibration”.
Note:
The sensor’s internal criteria for a successful product calibration are:
the sensor is currently in an environment corresponding to the pH ARC Sensors measurement range.
the manually assigned pH value does not deviate more than 2 pH units from the value measured prior the product calibration