Hamilton Electronics EPHUM011 manual Free User Memory Space, 0001001, Freeuserspace

Page 65

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

Command: Serial number

Modbus address: 1312

Length: 8

Type: 3

Read

Parameter:

Text

 

 

 

 

Format:

Character

 

 

 

 

Value:

0001001

 

 

 

 

Figure 2.9.2.2: Example to read register 1312.

2.9.3 Free User Memory Space

These registers can be used to store any customer specific information in the sensor. There are different registers which can be read by everybody, but only specific operators can write them.

Start

Number

Reg1 … Reg8

Example of content

Modbus

Read

Write

register

of

(16 ASCII characters)

 

function

access

access

 

registers

 

 

code

 

 

1536

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1544

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1552

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1560

8

Free user space U/A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

U/A/S

1568

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1576

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1584

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1592

8

Free user space A/S

*FREE_USERSPACE*

3, 4, 16

U/A/S

A/S

1600

8

Measuring point

242111-0001001

3, 4, 16

U/A/S

S

1608

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1616

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1624

8

Free user space S

*FREE_USERSPACE*

3, 4, 16

U/A/S

S

1632

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1640

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1648

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1656

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1664

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1672

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1680

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1688

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1696

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1704

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1712

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1720

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1728

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1736

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1744

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

1752

8

Free user space others

*FREE_USERSPACE*

3, 4

U/A/S

none

Figure 2.9.3.1: Definition of registers containing user information.

An important register is 1600, as it is the description of the measuring point. The information of this register is displayed on the ARC View Handheld in order to identify individual sensors.

Attention:

The Free User Memory Space is located in a memory which allows in total max 10’000 write operations.

624300/01

page 65 / 67

Image 65
Contents PH ARC Sensors Important Notice Configuration of the Analog Interfaces Configuration of the Serial RS485 InterfaceConfiguration of the Measurement Sensor Identification and InformationEPHUM011 IntroductionVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 Reading / Setting Operator Level Operator levels and PasswordsGeneral 0x030x30 Changing Passwords for Operator Level0x0C 18111978 0x0B 18111978Device Address Configuration of the serial RS485 InterfaceReading and Writing the Device Address Reading the Device Address LimitsReading the Baud Rate Limits Baud RateReading and Writing the Baud Rate Available Analog Interfaces Configuration of the Analog InterfacesAvailable Analog Interface Modes 0x07Description of the Analog Interfaces 1 Selection of an Analog Interface ModeMA interface #1 MA interface #20x01 Configuration of the 4-20 mA Interface0x21 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Defining the Measurement Values for 4, 12 and 20 mA Output Reading the Selected Physical Unit for Analog Interface0x001000 PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial Description of PMC1 Primary Measurement Channel 1 pHSelecting the Physical Unit for PMC1 0x2010000x200000 175.9922 0x00 954.6541 Reading the measurement value of PMC10x1000 02503 0x00 Description of PMC6 Primary Measurement Channel 6 TemperatureSelecting the Physical Unit for PMC6 0x06Definition of the Measurement Status for PMC1 / PMC6 Input of an Externally Measured TemperatureReading the measurement value of PMC6 0x04 24.35834 0x00 130Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 Selecting the Physical Unit and Writing the Value for PA12 3 PA12 Moving Average RDescription of PA12 Moving Average R Reading all Values for PA12 Available Calibration Points CalibrationCalibration Points 1 and 2 Standard Calibration Definitions of Calibration PointsCP1 CP2CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration StatusReading the Calibration Status of CP6 Product Calibration 0x00000080 0x000010000x00000180 0x00001000 0x08000000 0x00001000 0x01000000 0x000010000x0A000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 0x00000004 24.35184 23.78 Currently active Calibration Parameters part0x00000004 24.37691 16.45 0x00000004 29.93368 379.5167006804 099747 296.6901 003401 179.927 297.137810469 298.3302 607782 59.47631 298.15Special Commands for Calibration with Visical Available Sets of Calibration Standards Calibration StandardsLSB 0x00000001 Set 0x0000003F180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersDefinition of Warnings Currently Active WarningsHistory of Warnings History of Errors ErrorsCurrently Active Errors Definition of Errors CIP Reading Definition of SIP and CIPSIP 120 130 Reading the Sensor’s Quality IndicatorGeneral Information Sensor Identification and InformationSensor Identification EPHFI010Freeuserspace Free User Memory Space0001001 Recall Sensor’s Factory Settings System CommandsHello World 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011