Hamilton Electronics EPHUM011 Reading the Selected Physical Unit for Analog Interface, 0x001000

Page 13

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.4.5.5Reading the Selected Physical Unit for Analog Interface

Start

Number of

Reg1 / Reg2

Modbus

Read

Write

register

registers

 

function code

access

access

4376

2

Selected physical unit of AO1

3, 4

U/A/S

none

 

 

(see chapter 2.5.1)

 

 

 

4504

2

Selected physical unit of AO2

3, 4

U/A/S

none

 

 

(see chapter 2.5.1)

 

 

 

Figure 2.4.5.5.1: Definition of register 4376 / 4504.

Command: Avail unit AO1

Modbus address: 4376

Length: 2

Type: 3

Read

Parameter:

Available unit

 

 

 

 

 

 

Format:

Hex

 

 

 

 

 

 

Value:

0x001000

 

 

 

 

 

 

Figure 2.4.5.5.2: Example to read the selected unit of the selected PMC of AO1. The value returned is “0x001000”, accordingly, the unit is pH. The physical unit for PMC is defined in Reg. 2090 or 2410 and applies automatically for 4-20 mA output.

2.4.5.6Defining the Measurement Values for 4, 12 and 20 mA Output

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus

Read

Write

register

registers

 

 

 

function

access

access

 

 

 

 

 

code

 

 

4378

6

Measurement

Measurement

Measurement

3, 4, 16

U/A/S

S

 

 

value at

value at

value at

 

 

 

 

 

Min Output

Max Output

Mid Output

 

 

 

 

 

Current (4 mA)

Current (20

Current (12

 

 

 

 

 

for AO1

mA) for AO1

mA) for AO1

 

 

 

4506

6

Measurement

Measurement

Measurement

3, 4, 16

U/A/S

S

 

 

value at

value at

value at

 

 

 

 

 

Min Output

Max Output

Mid Output

 

 

 

 

 

Current (4 mA)

Current (20

Current (12

 

 

 

 

 

for AO2

mA) for AO2

mA) for AO2

 

 

 

Figure 2.4.5.6.1: Definition of register 4378 / 4506.

Command: MinMaxMid value AO1

Modbus address: 4378

Length: 6

Type: 16

Write

Parameter:

Min value

Max value

Mid value

 

 

Format:

Float

Float

Float

 

 

 

Value:

3

10

7

 

 

 

Figure 2.4.5.6.2: Example to set the min value to 3 (for 4 mA), the max value to 10 (for 20 mA) and the mid value to 7 (for 12 mA). The corresponding physical unit can be read in register 4376 / 4504 and in 2090 / 2410.

Note:

Mid current must always be defined. However, in linear output mode, the mid current value has no physical meaning and will not affect the 4-20 mA output.

624300/01

page 13 / 67

Image 13
Contents PH ARC Sensors Important Notice Configuration of the Analog Interfaces Configuration of the Serial RS485 InterfaceConfiguration of the Measurement Sensor Identification and InformationVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 IntroductionEPHUM011 Reading / Setting Operator Level Operator levels and PasswordsGeneral 0x030x30 Changing Passwords for Operator Level0x0C 18111978 0x0B 18111978Device Address Configuration of the serial RS485 InterfaceReading and Writing the Device Address Reading the Device Address LimitsReading and Writing the Baud Rate Baud RateReading the Baud Rate Limits Available Analog Interfaces Configuration of the Analog InterfacesAvailable Analog Interface Modes 0x07 Description of the Analog Interfaces 1 Selection of an Analog Interface Mode MA interface #1 MA interface #20x21 Configuration of the 4-20 mA Interface0x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 0x001000 Reading the Selected Physical Unit for Analog InterfaceDefining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial Description of PMC1 Primary Measurement Channel 1 pHSelecting the Physical Unit for PMC1 0x2010000x1000 02503 0x00 Reading the measurement value of PMC10x200000 175.9922 0x00 954.6541 Description of PMC6 Primary Measurement Channel 6 TemperatureSelecting the Physical Unit for PMC6 0x06Definition of the Measurement Status for PMC1 / PMC6 Input of an Externally Measured TemperatureReading the measurement value of PMC6 0x04 24.35834 0x00 130Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 Description of PA12 Moving Average R 3 PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Available Calibration Points CalibrationCalibration Points 1 and 2 Standard Calibration Definitions of Calibration PointsCP1 CP2CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration Status0x00000180 0x00001000 0x00000080 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x08000000 0x00001000 0x01000000 0x000010000x0A000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 0x00000004 24.35184 23.78 Currently active Calibration Parameters part0x00000004 24.37691 16.45 0x00000004 29.93368 379.5167006804 099747 296.6901 003401 179.927 297.137810469 298.3302 607782 59.47631 298.15Special Commands for Calibration with Visical Available Sets of Calibration Standards Calibration StandardsLSB 0x00000001 Set 0x0000003F180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersHistory of Warnings Currently Active WarningsDefinition of Warnings Currently Active Errors ErrorsHistory of Errors Definition of Errors SIP Reading Definition of SIP and CIPCIP 120 130 Reading the Sensor’s Quality IndicatorGeneral Information Sensor Identification and InformationSensor Identification EPHFI0100001001 Free User Memory SpaceFreeuserspace Recall Sensor’s Factory Settings System CommandsHello World 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011