pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
Attention:
When assigning measurement values to
The PMC you have mapped to AO1 / AO2 (register 4364 / 4492)
The physical unit currently in use for the selected PMC (register 2090 for PMC1 (pH) and
register 2410 for PMC6 (temperature).
Therefore, when the operator redefines one of the register 4364 / 4492, 2090 / 2410, the definitions of the register 4378 / 4506 should be reviewed. If not, the current output at the
Note:
The physical unit of the analog output corresponds always to the unit that is set for the selected PMC (register 2090 for PMC1 or register 2410 for PMC6). Accordingly, not only the pH value is selectable at the
Example:
Register 4364 is set to 1 (PMC1 is mapped to AO1).
Register 2090 is set to 0x1000 (the unit “pH”is assigned to PMC1). Register 4378 is set to 2 and 10 (4 mA = pH 2, 20 mA = pH 10).
The sensor reads currently pH 4, the output at the
The operator now
This will generate an interface warning
2.4.5.7Defining a Constant Current Output for Testing
Note:
For constant current output, the AO1 / AO2 must be set to analog interface mode 0x01:
Start | Number of | Reg1 / Reg2 | Modbus | Read | Write |
register | registers |
| function code | access | access |
4384 | 2 | Constant current output value for | 3, 4, 16 | U/A/S | S |
|
| AO1 [mA] |
|
|
|
4512 | 2 | Constant current output value for | 3, 4, 16 | U/A/S | S |
|
| AO2 [mA] |
|
|
|
Figure 2.4.5.7.1: Definition of register 4384 / 4512.
Command: Fixed value AO1 | Modbus address: 4384 | Length: 2 | Type: 3 | Read | |||
Parameter: | Fixed value [mA] |
|
|
|
|
|
|
Format: | Float |
|
|
|
|
|
|
Value: | 10 |
|
|
|
|
|
|
Figure 2.4.5.7.2: Example to read the constant current output in mode 0x01 for AO1. It is set to 10 mA.
624300/01 | page 15 / 67 |