Hamilton Electronics EPHUM011 manual Calibration Procedure

Page 35

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.3Calibration Procedure

2.7.3.1Calibration at CP1 and CP2 (Standard Calibration)

The ARC Sensor family has a unique calibration routine. When initiating the calibration, the data set of the sensor is automatically traced back within the last 3 minutes and a decision is made immediately if the calibration is successful or not. The operator therefore gets an immediate result. The criteria for a successful calibration are:

the stability of pH value and temperature over the last 3 minutes (see register 5128)

the currently measured pH value fits to one of the calibration standards defined in the selected set of calibration standards

the limits of slope and offset at pH 7 have to be met

 

 

 

Standard Calibration at 2 calibration points

 

 

 

 

200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ref [mV]

150

 

 

 

 

CP1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

vs.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

glass

50

 

 

 

 

 

 

 

CP2

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

potential

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-50

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

electrical

-100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-150

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-200

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

1

2

3

4

5

6

7

8

9

10

11

12

13

14

 

 

 

 

 

 

 

 

pH value

 

 

 

 

 

 

Figure 2.7.3.1.1: Standard Calibration using CP1 and CP2.

CP1 and CP2 define a linear relationship between the electrical potential and the pH value. This linear calibration function is defined by an offset at pH 7 [mV] and a slope [mV/pH]. These two values are stored in register 5448.

Note: the pH value of CP1 can be lower or higher than the pH value of CP2. However, the difference in pH between CP1 and CP2 must always be greater than 1 pH unit.

Notes:

In order to perform a standard calibration at CP1 and CP2, it is necessary to use commercially available calibration standards. The operator is restricted to use those standards that are defined in the six sets of calibration standards (see chapter 2.7.11).

There are two ways of performing a standard calibration:

- standard calibration with automatic recognition of the calibration standard:

in this case, the sensor decides on itself in what calibration standard it is immersed. The criteria to decide on is the electrical potential measured. The sensor checks the list of calibration standards that are available for automatic recognition (see chapter 2.7.11)

- standard calibration with manual selection of the calibration standard:

the operator is selecting the calibration standard in which the sensor is immersed.

Factory calibration is pH 4 at CP1 and pH 7 at CP 2.

624300/01

page 35 / 67

Image 35
Contents PH ARC Sensors Important Notice Sensor Identification and Information Configuration of the Serial RS485 InterfaceConfiguration of the Analog Interfaces Configuration of the MeasurementEPHUM011 IntroductionVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 0x03 Operator levels and PasswordsReading / Setting Operator Level General0x0B 18111978 Changing Passwords for Operator Level0x30 0x0C 18111978Reading the Device Address Limits Configuration of the serial RS485 InterfaceDevice Address Reading and Writing the Device AddressReading the Baud Rate Limits Baud RateReading and Writing the Baud Rate 0x07 Configuration of the Analog InterfacesAvailable Analog Interfaces Available Analog Interface ModesMA interface #2 Selection of an Analog Interface ModeDescription of the Analog Interfaces 1 MA interface #10x01 Configuration of the 4-20 mA Interface0x21 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Defining the Measurement Values for 4, 12 and 20 mA Output Reading the Selected Physical Unit for Analog Interface0x001000 PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial 0x201000 Primary Measurement Channel 1 pHDescription of PMC1 Selecting the Physical Unit for PMC10x200000 175.9922 0x00 954.6541 Reading the measurement value of PMC10x1000 02503 0x00 0x06 Primary Measurement Channel 6 TemperatureDescription of PMC6 Selecting the Physical Unit for PMC60x04 24.35834 0x00 130 Input of an Externally Measured TemperatureDefinition of the Measurement Status for PMC1 / PMC6 Reading the measurement value of PMC6Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 Selecting the Physical Unit and Writing the Value for PA12 3 PA12 Moving Average RDescription of PA12 Moving Average R Reading all Values for PA12 Available Calibration Points CalibrationCP2 Definitions of Calibration PointsCalibration Points 1 and 2 Standard Calibration CP1CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration StatusReading the Calibration Status of CP6 Product Calibration 0x00000080 0x000010000x00000180 0x00001000 0x14000000 0x00001000 0x01000000 0x000010000x08000000 0x00001000 0x0A000000 0x000010000x10000000 0x00001000 0x00000004 29.93368 379.5167 Currently active Calibration Parameters part0x00000004 24.35184 23.78 0x00000004 24.37691 16.45607782 59.47631 298.15 003401 179.927 297.1378006804 099747 296.6901 10469 298.3302Special Commands for Calibration with Visical 0x0000003F Calibration StandardsAvailable Sets of Calibration Standards LSB 0x00000001 Set180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersDefinition of Warnings Currently Active WarningsHistory of Warnings History of Errors ErrorsCurrently Active Errors Definition of Errors CIP Reading Definition of SIP and CIPSIP 120 130 Reading the Sensor’s Quality IndicatorEPHFI010 Sensor Identification and InformationGeneral Information Sensor IdentificationFreeuserspace Free User Memory Space0001001 732255 System CommandsRecall Sensor’s Factory Settings Hello WorldPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011