pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.2 Definitions of Calibration Points

2.7.2.1Calibration Points 1 and 2 (Standard Calibration)

The limits for the calibration point 1 are defined in register 5152, for calibration point 2 in register 5184.

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg5 / Reg6

Modbus

Read

Write

register

registers

 

 

 

function code

access

access

5152

6

Physical unit

Min value for

Max value for

3, 4

U/A/S

none

 

 

currently

CP1 (in the

CP1 (in the

 

 

 

 

 

active for

physical unit

physical unit

 

 

 

 

 

CP1

as defined in

as defined in

 

 

 

 

 

 

Reg1 and 2)

Reg1 and 2)

 

 

 

5184

6

Physical unit

Min value for

Max value for

3, 4

U/A/S

none

 

 

currently

CP2 (in the

CP2 (in the

 

 

 

 

 

active for

physical unit

physical unit

 

 

 

 

 

CP2

as defined in

as defined in

 

 

 

 

 

 

Reg1 and 2)

Reg1 and 2)

 

 

 

Figure 2.7.2.1.1: Definition of register 5152 for CP1 and 5184 for CP2.

Attention:

The only physical unit available for calibration is pH ! The physical unit defined in 5152, 5184 and 5312 for CP1, CP2 and CP6 is NOT linked to the physical unit defined for PMC1 in register 2090.

When performing a calibration while having the physical unit set to “mV”, the calibration status will report “CP1: incorrect measurement unit”or “CP2: incorrect measurement unit”(see figure 2.7.4.1.1).

Command: Calibration limits CP1

 

Modbus address: 5152

Length: 6

Type: 3

Read

Parameter:

Unit

Min value

Max value

 

 

Format:

Hex

Float

Float

 

 

 

Value:

0x01000

0

 

0

 

 

 

Figure 2.7.2.1.2: Example to read the limits of CP1. Currently active physical unit is pH (0x01000). The min and max values are both 0, indicating, that calibration at CP1 can be performed only using defined calibration standards having discrete pH values.

Command: Calibration limits CP2

 

Modbus address: 5184

Length: 6

Type: 3

Read

Parameter:

Unit

Min value

Max value

 

 

Format:

Hex

Float

Float

 

 

 

Value:

0x01000

0

 

0

 

 

 

Figure 2.7.2.1.3: Example to read the limits of CP2. The active physical unit is is pH (0x01000). The min and max values are both 0, indicating, that calibration at CP2 can be performed only using defined calibration standards having discrete pH values.

When initiating the calibration at CP1 and CP2, the measured pH and temperature have to be stable for at least 3 minutes. The stability criteria are defined in register 5128:

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Modbus

Read

Write

register

registers

(Float)

(Float)

function code

access

access

5128

4

Max. Drift PMC1

Max. Drift PMC6

3, 4, 16

U/A/S

S

 

 

pH

Temperature

 

 

 

 

 

[pH/min]

[K/min]

 

 

 

Figure 2.7.2.1.4: Definition of register 5128.

624300/01

page 33 / 67

Image 33
Contents PH ARC Sensors Important Notice Configuration of the Analog Interfaces Configuration of the Serial RS485 InterfaceConfiguration of the Measurement Sensor Identification and InformationIntroduction Visiferm do Modbus RTU Programmer’s Manualp/n 624179/01EPHUM011 Reading / Setting Operator Level Operator levels and PasswordsGeneral 0x030x30 Changing Passwords for Operator Level0x0C 18111978 0x0B 18111978Device Address Configuration of the serial RS485 InterfaceReading and Writing the Device Address Reading the Device Address LimitsBaud Rate Reading and Writing the Baud RateReading the Baud Rate Limits Available Analog Interfaces Configuration of the Analog InterfacesAvailable Analog Interface Modes 0x07Description of the Analog Interfaces 1 Selection of an Analog Interface ModeMA interface #1 MA interface #2Configuration of the 4-20 mA Interface 0x210x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Reading the Selected Physical Unit for Analog Interface 0x001000Defining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 99186 99742 Reading the Internally Measured Output CurrentDefinition of Measurement Channels and Physical Units MeasurementSpecial Description of PMC1 Primary Measurement Channel 1 pHSelecting the Physical Unit for PMC1 0x201000Reading the measurement value of PMC1 0x1000 02503 0x000x200000 175.9922 0x00 954.6541 Description of PMC6 Primary Measurement Channel 6 TemperatureSelecting the Physical Unit for PMC6 0x06Definition of the Measurement Status for PMC1 / PMC6 Input of an Externally Measured TemperatureReading the measurement value of PMC6 0x04 24.35834 0x00 130Description of SMC Secondary Measurement Channels0x8000 247.56 Reading the measurement value of SMCAvailable Parameters Configuration of the MeasurementEffect of Moving Average n=1 vs n=16 2 PA9 Moving AverageSelecting the Physical Unit and Writing the Value for PA9 Description of PA9 Moving AverageReading all Values for PA9 3 PA12 Moving Average R Description of PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Available Calibration Points CalibrationCalibration Points 1 and 2 Standard Calibration Definitions of Calibration PointsCP1 CP2CP6 Calibration Point 6 Product CalibrationCalibration at CP1 and CP2 Standard Calibration Calibration ProcedurePH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 Calibration at CP6 Product Calibration 100200 Product calibration Initial measurement Definition of the commands for product calibrationProduct calibration Assignment Product calibration CancelProduct calibration Restore product calibration Product calibration Restore standard calibrationReading the Calibration Status of CP1 and CP2 Reading the Calibration Status0x00000080 0x00001000 0x00000180 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x08000000 0x00001000 0x01000000 0x000010000x0A000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 0x00000004 24.35184 23.78 Currently active Calibration Parameters part0x00000004 24.37691 16.45 0x00000004 29.93368 379.5167006804 099747 296.6901 003401 179.927 297.137810469 298.3302 607782 59.47631 298.15Special Commands for Calibration with Visical Available Sets of Calibration Standards Calibration StandardsLSB 0x00000001 Set 0x0000003F180 Definitions for Individual Sets of Calibration Standards0x0FFF0FFF 0x05540FFF 0x02480FFFNist Standard DINMettler Toledo Radiometer Sensor Status Temperature Ranges168.3667 Operating Hours and CountersCurrently Active Warnings History of WarningsDefinition of Warnings Errors Currently Active ErrorsHistory of Errors Definition of Errors Reading Definition of SIP and CIP SIPCIP 120 130 Reading the Sensor’s Quality IndicatorGeneral Information Sensor Identification and InformationSensor Identification EPHFI010Free User Memory Space 0001001Freeuserspace Recall Sensor’s Factory Settings System CommandsHello World 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011