pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.5 Currently active Calibration Parameters part 1

In registers 5164 (CP1), 5196 (CP2) and 5324 (CP6) the currently active calibration parameters part 1 are stored. These registers contain the values for temperature, number of calibrations and operating hour upon calibration.

Start

Number

Reg1 / Reg2

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

of

 

Reg4

Reg6

Reg8

function

access

access

 

registers

 

 

 

 

code

 

 

5164

8

Unit of

Value of

Number of

Operating

3, 4

U/A/S

none

 

 

temperature for

temperature

calibrations

hour for

 

 

 

 

 

CP1 (bitwise

of CP1

at CP1

CP1

 

 

 

 

 

defined)

 

 

 

 

 

 

5196

8

Unit of

Value of

Number of

Operating

3, 4

U/A/S

none

 

 

temperature for

temperature

calibrations

hour for

 

 

 

 

 

CP2 (bitwise

of CP2

at CP2

CP2

 

 

 

 

 

defined)

 

 

 

 

 

 

5324

8

Unit of

Value of

Number of

Operating

3, 4

U/A/S

none

 

 

temperature for

temperature

calibrations

hour for

 

 

 

 

 

CP6 (bitwise

of CP6

at CP6

CP6

 

 

 

 

 

defined)

 

 

 

 

 

 

Figure 2.7.5.1: Definition of register 5164 for CP1, 5196 for CP2 and 5324 for CP6.

Command: Calibration CP1 values

Modbus address: 5164

 

Length: 8

Type: 3

Read

Parameter:

Unit of temperature

Temperature

Number

of cali

Operating hour

Format:

Hex

Float

Decimal

 

Float

 

Value:

0x00000004

24.35184

6

 

 

23.78

 

Figure 2.7.5.2: Example to read the calibration values for CP1. The physical unit is °C (0x00000004), the temperature is 24.35184 °C, the number of calibrations at CP1 is 6 and the operating hour is 23.78 h.

Command: Calibration CP2 values

Modbus address: 5196

 

Length: 8

Type: 3

Read

Parameter:

Unit of temperature

Temperature

Number

of cali

Operating hour

Format:

Hex

Float

Decimal

 

Float

 

Value:

0x00000004

24.37691

5

 

 

16.45

 

Figure 2.7.5.3: Example to read the calibration values for CP2. The physical unit is °C (0x00000004), the temperature is 24.37691 °C, the number of calibrations at CP1 is 5 and the operating hour is 16.45 h.

Command: Calibration CP6 values

Modbus address: 5324

 

Length: 8

Type: 3

Read

Parameter:

Unit of temperature

Temperature

Number

of cali

Operating hour

Format:

Hex

Float

Decimal

 

Float

 

Value:

0x00000004

29.93368

12

 

 

379.5167

 

Figure 2.7.5.4: Example to read the calibration values 1 for CP6. The physical unit is °C (4), the temperature is 29.93 (°C), the number of calibrations at CP1 is 12 and the operating hour is

379.51(h).

2.7.6Currently active Calibration Parameters part 2

Registers 5172 (CP1), 5204 (CP2) and 5332 (CP6) are not defined for pH ARC Sensors, as they document atmospheric pressure and salinity used for VISIFERM DO / VISIFERM DO ARC Sensors only.

624300/01

page 48 / 67

Image 48
Contents PH ARC Sensors Important Notice Configuration of the Serial RS485 Interface Configuration of the Analog InterfacesConfiguration of the Measurement Sensor Identification and InformationIntroduction Visiferm do Modbus RTU Programmer’s Manualp/n 624179/01EPHUM011 Operator levels and Passwords Reading / Setting Operator LevelGeneral 0x03Changing Passwords for Operator Level 0x300x0C 18111978 0x0B 18111978Configuration of the serial RS485 Interface Device AddressReading and Writing the Device Address Reading the Device Address LimitsBaud Rate Reading and Writing the Baud RateReading the Baud Rate Limits Configuration of the Analog Interfaces Available Analog InterfacesAvailable Analog Interface Modes 0x07Selection of an Analog Interface Mode Description of the Analog Interfaces 1MA interface #1 MA interface #2Configuration of the 4-20 mA Interface 0x210x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Reading the Selected Physical Unit for Analog Interface 0x001000Defining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 Reading the Internally Measured Output Current 99186 99742Measurement Definition of Measurement Channels and Physical UnitsSpecial Primary Measurement Channel 1 pH Description of PMC1Selecting the Physical Unit for PMC1 0x201000Reading the measurement value of PMC1 0x1000 02503 0x000x200000 175.9922 0x00 954.6541 Primary Measurement Channel 6 Temperature Description of PMC6Selecting the Physical Unit for PMC6 0x06Input of an Externally Measured Temperature Definition of the Measurement Status for PMC1 / PMC6Reading the measurement value of PMC6 0x04 24.35834 0x00 130Secondary Measurement Channels Description of SMCReading the measurement value of SMC 0x8000 247.56Configuration of the Measurement Available Parameters2 PA9 Moving Average Effect of Moving Average n=1 vs n=16Description of PA9 Moving Average Selecting the Physical Unit and Writing the Value for PA9Reading all Values for PA9 3 PA12 Moving Average R Description of PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Calibration Available Calibration PointsDefinitions of Calibration Points Calibration Points 1 and 2 Standard CalibrationCP1 CP2Calibration Point 6 Product Calibration CP6Calibration Procedure Calibration at CP1 and CP2 Standard CalibrationPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 100 Calibration at CP6 Product Calibration200 Definition of the commands for product calibration Product calibration Initial measurementProduct calibration Cancel Product calibration AssignmentProduct calibration Restore standard calibration Product calibration Restore product calibrationReading the Calibration Status Reading the Calibration Status of CP1 and CP20x00000080 0x00001000 0x00000180 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x01000000 0x00001000 0x08000000 0x000010000x0A000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 Currently active Calibration Parameters part 0x00000004 24.35184 23.780x00000004 24.37691 16.45 0x00000004 29.93368 379.5167003401 179.927 297.1378 006804 099747 296.690110469 298.3302 607782 59.47631 298.15Special Commands for Calibration with Visical Calibration Standards Available Sets of Calibration StandardsLSB 0x00000001 Set 0x0000003FDefinitions for Individual Sets of Calibration Standards 1800x0FFF0FFF 0x02480FFF 0x05540FFFDIN Nist StandardMettler Toledo Radiometer Temperature Ranges Sensor StatusOperating Hours and Counters 168.3667Currently Active Warnings History of WarningsDefinition of Warnings Errors Currently Active ErrorsHistory of Errors Definition of Errors Reading Definition of SIP and CIP SIPCIP Reading the Sensor’s Quality Indicator 120 130Sensor Identification and Information General InformationSensor Identification EPHFI010Free User Memory Space 0001001Freeuserspace System Commands Recall Sensor’s Factory SettingsHello World 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011