pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.8.2 Operating Hours and Counters

In register 4676 are given:

total operating hours

operating hours above max measurement temperature (see chapter 2.8.1)

the operating hours above max operating temperature (see chapter 2.8.1)

In register 4682 are stored:

number of power ups

number of watchdog resets

number of writing cycles to the sensor’s flash memory

In register 4688 are given:

number of sterilizations in place (SIP) (see chapter 2.8.5)

number of cleanings in place (CIP) (see chapter 2.8.5)

Start

Number of

Reg1 / Reg2

Reg3 / Reg4

Reg3 / Reg4

Modbus

Read

Write

register

registers

 

 

 

function code

access

access

4676

6

Operating

Operating

Operating

3, 4

U/A/S

none

 

 

hours

hours above

hours above

 

 

 

 

 

[h]

max

max

 

 

 

 

 

 

measurement

operating

 

 

 

 

 

 

temperature

temperature

 

 

 

 

 

 

[h]

[h]

 

 

 

4682

6

Number of

Number of

Number of

3, 4

U/A/S

none

 

 

Power ups

Watchdog

Writing

 

 

 

 

 

 

resets

cycles to

 

 

 

 

 

 

 

flash memory

 

 

 

4688

4

Number of

Number of

-

3, 4

U/A/S

none

 

 

SIP cycles

CIP cycles

 

 

 

 

Figure 2.8.2.1: Definition of register 4676, 4682 and 4688.

Command: Operating hours

 

Modbus address: 4676

Length: 6

Type: 3

Read

Parameter:

Operating hours [h]

 

Operating hours

Operating hours

 

 

 

 

above max

above max

 

 

 

 

measurement

operating

 

 

 

 

temperature

temperature

 

 

 

 

[h]

[h]

 

 

 

Format:

Float

Float

Float

 

 

 

Value:

168.3667

0

 

0

 

 

 

Figure 2.8.2.2: Example to read the total operating hours, the operating hours above the max measurement temperature and the operating hours above the max operating temperature.

Command: Power & watchdog

 

Modbus address: 4682

 

Length: 6

Type: 3

Read

Parameter:

Number of Power

Number of

Number

of Writing

 

 

 

ups

Watchdog resets

cycles to flash

 

 

 

 

 

 

memory

 

 

 

Format:

Decimal

Decimal

Decimal

 

 

 

Value:

34

1

 

16

 

 

 

 

Figure 2.8.2.3: Example to read the number of power ups, the number of watchdog resets and the number of writing cycles to flash memory.

Command: SIP & CIP

 

Modbus address: 4688

Length: 4

Type: 3

Read

Parameter:

SIP cycles

CIP cycles

 

 

 

 

Format:

Decimal

Decimal

 

 

 

 

Value:

0

0

 

 

 

 

 

Figure 2.8.2.4: Example to read the number of SIP cycles and the number of CIP cycles. For the definition of SIP and CIP cycles see chapter 2.8.5.

624300/01

page 58 / 67

Image 58
Contents PH ARC Sensors Important Notice Configuration of the Measurement Configuration of the Serial RS485 InterfaceConfiguration of the Analog Interfaces Sensor Identification and InformationVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 IntroductionEPHUM011 General Operator levels and PasswordsReading / Setting Operator Level 0x030x0C 18111978 Changing Passwords for Operator Level0x30 0x0B 18111978Reading and Writing the Device Address Configuration of the serial RS485 InterfaceDevice Address Reading the Device Address LimitsReading and Writing the Baud Rate Baud RateReading the Baud Rate Limits Available Analog Interface Modes Configuration of the Analog InterfacesAvailable Analog Interfaces 0x07MA interface #1 Selection of an Analog Interface ModeDescription of the Analog Interfaces 1 MA interface #20x21 Configuration of the 4-20 mA Interface0x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 0x001000 Reading the Selected Physical Unit for Analog InterfaceDefining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 Reading the Internally Measured Output Current 99186 99742Measurement Definition of Measurement Channels and Physical UnitsSpecial Selecting the Physical Unit for PMC1 Primary Measurement Channel 1 pHDescription of PMC1 0x2010000x1000 02503 0x00 Reading the measurement value of PMC10x200000 175.9922 0x00 954.6541 Selecting the Physical Unit for PMC6 Primary Measurement Channel 6 TemperatureDescription of PMC6 0x06Reading the measurement value of PMC6 Input of an Externally Measured TemperatureDefinition of the Measurement Status for PMC1 / PMC6 0x04 24.35834 0x00 130Secondary Measurement Channels Description of SMCReading the measurement value of SMC 0x8000 247.56Configuration of the Measurement Available Parameters2 PA9 Moving Average Effect of Moving Average n=1 vs n=16Description of PA9 Moving Average Selecting the Physical Unit and Writing the Value for PA9Reading all Values for PA9 Description of PA12 Moving Average R 3 PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Calibration Available Calibration PointsCP1 Definitions of Calibration PointsCalibration Points 1 and 2 Standard Calibration CP2Calibration Point 6 Product Calibration CP6Calibration Procedure Calibration at CP1 and CP2 Standard CalibrationPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 100 Calibration at CP6 Product Calibration200 Definition of the commands for product calibration Product calibration Initial measurementProduct calibration Cancel Product calibration AssignmentProduct calibration Restore standard calibration Product calibration Restore product calibrationReading the Calibration Status Reading the Calibration Status of CP1 and CP20x00000180 0x00001000 0x00000080 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x0A000000 0x00001000 0x01000000 0x000010000x08000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 0x00000004 24.37691 16.45 Currently active Calibration Parameters part0x00000004 24.35184 23.78 0x00000004 29.93368 379.516710469 298.3302 003401 179.927 297.1378006804 099747 296.6901 607782 59.47631 298.15Special Commands for Calibration with Visical LSB 0x00000001 Set Calibration StandardsAvailable Sets of Calibration Standards 0x0000003FDefinitions for Individual Sets of Calibration Standards 1800x0FFF0FFF 0x02480FFF 0x05540FFFDIN Nist StandardMettler Toledo Radiometer Temperature Ranges Sensor StatusOperating Hours and Counters 168.3667History of Warnings Currently Active WarningsDefinition of Warnings Currently Active Errors ErrorsHistory of Errors Definition of Errors SIP Reading Definition of SIP and CIPCIP Reading the Sensor’s Quality Indicator 120 130Sensor Identification Sensor Identification and InformationGeneral Information EPHFI0100001001 Free User Memory SpaceFreeuserspace Hello World System CommandsRecall Sensor’s Factory Settings 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011