pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
2.4Configuration of the Analog Interfaces
2.4.1 Available Analog Interfaces
A pH ARC Sensor has two individual physical analog interfaces that have identical functionalities, but can be configured independently from each other.
Analog Output Interface 1 (AO1)
Analog Output Interface 2 (AO2)
The number of analog interfaces is defined in register 4320.
Start | Number of | Reg1 / Reg2 | Modbus | Read | Write |
register | registers |
| function code | access | access |
4320 | 2 | Available analog interfaces | 3, 4 | U/A/S | none |
Figure 2.4.1.1: Definition of register 4320.
Command: Avail analog interfaces | Modbus address: 4320 | Length: 2 | Type: 3 | Read | |||
Parameter: | Available analog |
|
|
|
|
|
|
| interfaces |
|
|
|
|
|
|
Format: | Hex |
|
|
|
|
|
|
Value: | 0x03 |
|
|
|
|
|
|
Figure 2.4.1.2: Example to read the available analog interfaces. The answer is “0x03”meaning that there exists an Analog Interface 1 (AO1) and an Analog Interface 2 (AO2).
2.4.2 Available Analog Interface Modes
With register 4322, the available analog interface modes for AO1 and AO2 are defined
Start |
| Number of |
| Reg1 / |
| Reg3 / | Reg5 / | Reg7 / |
| Modbus | Read |
| Write | |
register |
| registers |
| Reg2 |
| Reg4 | Reg6 | Reg8 |
| function | access |
| access | |
|
|
|
|
|
|
|
|
|
|
| code |
|
|
|
4322 | 8 |
| Available |
| Available | reserved | reserved |
| 3,4 | U/A/S |
| none | ||
|
|
|
| Analog |
| Analog |
|
|
|
|
|
|
| |
|
|
|
| Interface |
| Interface |
|
|
|
|
|
|
| |
|
|
|
| Modes for | Modes |
|
|
|
|
|
|
| ||
|
|
|
| AO1 |
| for AO2 |
|
|
|
|
|
|
| |
Figure 2.4.2.1: Definition of register 4322. It defines the analog interface modes available for AO1 and | ||||||||||||||
AO2. The analog interface modes are described in Figure 2.4.2.2. |
|
|
|
| ||||||||||
|
|
|
|
|
|
|
|
|
|
|
| |||
Code |
| Analog Interface |
| Description |
|
|
|
|
|
|
| |||
(Hex) |
| Mode |
|
|
|
|
|
|
|
|
|
|
| |
0x00 |
|
| Analog interface deactivated |
|
|
|
|
|
| |||||
0x01 |
|
|
| Set to a constant output value for current loop testing |
|
|
| |||||||
0x02 |
|
| Linear output of measurement (PMC1 / 6) |
|
|
|
| |||||||
0x04 |
|
| Bilinear output of measurement (PMC1 / 6) |
|
|
|
|
Figure 2.4.2.2: Definition of the analog interface modes, valid for both AO1 and AO2.
Command: Analog Interface Modes | Modbus address: 4322 | Length: 8 | Type: 3 | Read | ||
Parameter: | Available Analog | Available Analog | reserved |
| reserved |
|
| Interface Modes for | Interface Modes for |
|
|
|
|
| AO1 | AO2 |
|
|
|
|
Format: | Hex | Hex | Hex |
| Hex |
|
Value: | 0x07 | 0x07 | 0x0 |
| 0x0 |
|
Figure 2.4.2.3: Example to read register 4322: all modes defined in figure 2.4.2.2 are available for both AO1 and AO2.