pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)
Command: Read calibration stability | Modbus address: 5128 | Length: 4 | Type: 3 | Read | ||
Parameter: | Max drift pH |
| Max drift Temp |
|
|
|
| [pH/min] |
| [K/min] |
|
|
|
Format: | Float |
| Float |
|
|
|
Value: | 0.1 |
| 0.5 |
|
|
|
Figure 2.7.2.1.5: | Example to read the | calibration stability. |
|
|
| |
|
|
|
|
| ||
Command: Set calibration stability | Modbus address: 5128 | Length: 4 | Type: 16 | Write | ||
Parameter: | Max drift pH |
| Max drift Temp |
|
|
|
| [pH/min] |
| [K/min] |
|
|
|
Format: | Float |
| Float |
|
|
|
Value: | 0.2 |
| 0.5 |
|
|
|
Figure 2.7.2.1.6: Example to set the calibration stability.
Attention:
The stability criteria defined in register 5128 is valid for CP1 and CP2 only, but NOT for CP6.
2.7.2.2Calibration Point 6 (Product Calibration)
The limits for calibration point 6 are given in register 5312.
Start | Number of | Reg1 / Reg2 | Reg3 / Reg4 | Reg5 / Reg6 | Modbus | Read | Write |
register | registers |
|
|
| function code | access | access |
5312 | 6 | Physical unit | Min value for | Max value for | 3, 4 | U/A/S | none |
|
| currently | CP6 (in the | CP6 (in the |
|
|
|
|
| active for | physical unit | physical unit |
|
|
|
|
| CP6 | as defined in | as defined in |
|
|
|
|
|
| Reg1 and 2) | Reg1 and 2) |
|
|
|
Figure 2.7.2.2.1: Definition of register 5312 for CP6.
Command: Calibration limits CP6 |
| Modbus address: 5312 | Length: 6 | Type: 3 | Read | ||
Parameter: | Unit | Min value | Max value |
|
| ||
Format: | Hex | Float | Float |
|
|
| |
Value: | 0x01000 | 0 |
| 14 |
|
|
|
Figure 2.7.2.2.2: Example to read the limits of CP6. The active physical unit is pH, the min value is pH 0 and the max value is pH 14.
Note: the definition of min and max is different than the one for CP1 / CP2, because CP6 can be set to any pH value.
624300/01 | page 34 / 67 |