Page 52

pH ARC Sensors Modbus RTU Programmer’s Manual (EPHUM011)

2.7.10.2 Definitions for Individual Sets of Calibration Standards

Once the operator has selected a set of calibration standards (register 9474) the register 9504 and following give all information on the selected set of calibration standards.

Start

Number of

Reg1 … Reg8

Modbus

Read

Write

register

registers

(16 ASCII characters)

function code

access

access

9504

8

Manufacturer of the selected set

3, 4

U/A/S

none

9512

8

Info 1 of the selected set

3, 4

U/A/S

none

9520

8

Info 2 of the selected set

3, 4

U/A/S

none

Figure 2.7.10.2.1: Definition of registers 9504 to 9520.

Command: manufacturer of set

Modbus address: 9504

Length: 8

Type: 3

Read

Parameter:

Text

 

 

 

 

Format:

Character

 

 

 

 

Value:

HAMILTON

 

 

 

 

Figure 2.7.10.2.2: Example to read the description (in this case, the set 0x01 is selected)

Within one set of calibration standards, a maximum of 12 calibration standards are available. In the following registers, the details of each calibration standard is given:

Start register

Description

 

 

 

 

 

 

 

9536

 

Nominal value and tolerance of

calibration standard 1

 

 

 

 

9552

 

Nominal value and tolerance of

calibration standard 2

 

 

 

 

9568

 

Nominal value and tolerance of

calibration standard 3

 

 

 

 

9584

 

Nominal value and tolerance of

calibration standard 4

 

 

 

 

9600

 

Nominal value and tolerance of

calibration standard 5

 

 

 

 

9616

 

Nominal value and tolerance of

calibration standard 6

 

 

 

 

9632

 

Nominal value and tolerance of

calibration standard 7

 

 

 

 

9648

 

Nominal value and tolerance of

calibration standard 8

 

 

 

 

9664

 

Nominal value and tolerance of

calibration standard 9

 

 

 

 

9680

 

Nominal value and tolerance of

calibration standard 10

 

 

 

 

9696

 

Nominal value and tolerance of

calibration standard 11

 

 

 

 

9712

 

Nominal value and tolerance of

calibration standard 12

 

 

 

 

Figure 2.7.10.2.3:

Definition for the register range from 9536 until 9720.

 

 

 

 

 

 

 

 

 

 

 

 

 

Start

Number

Reg1 /

Reg3 /

Reg5 /

Reg7 /

Modbus

Read

Write

register

of

Reg2

Reg4

Reg6

Reg8

function

access

access

 

registers

(Float)

(Float)

(Float)

(Float)

code

 

 

9536,

8

 

Calibr.

Tolerance

Nominal

Tolerance

3, 4

U/A/S

none

9552,

 

 

standard

of nominal

electrical

of electrical

 

 

 

 

 

 

nominal

value (±)

potential

potential

 

 

 

 

 

 

 

value

[pH]

[mV]

(±) [mV]

 

 

 

 

 

 

 

[pH]

 

 

 

 

 

 

 

Figure 2.7.10.2.4: Definition for registers 9536 until 9720.

Command: Read standard 4

Modbus address: 9584

Length: 8

Type: 3

Read

Parameter:

Nominal value

Tolerance

Nominal

electrical

Tolerance of

 

 

[pH]

(±) [pH]

potential

electrical potential

 

 

 

[mV]

 

(±) [mV]

 

Format:

Float

Float

Float

 

Float

 

Value:

4.01

0.02

180

 

80

 

Figure 2.7.10.2.5: Example to read the values for the calibration standard 4.

624300/01

page 52 / 67

Image 52
Contents PH ARC Sensors Important Notice Configuration of the Serial RS485 Interface Configuration of the Analog InterfacesConfiguration of the Measurement Sensor Identification and InformationVisiferm do Modbus RTU Programmer’s Manualp/n 624179/01 IntroductionEPHUM011 Operator levels and Passwords Reading / Setting Operator LevelGeneral 0x03Changing Passwords for Operator Level 0x300x0C 18111978 0x0B 18111978Configuration of the serial RS485 Interface Device AddressReading and Writing the Device Address Reading the Device Address LimitsReading and Writing the Baud Rate Baud RateReading the Baud Rate Limits Configuration of the Analog Interfaces Available Analog InterfacesAvailable Analog Interface Modes 0x07Selection of an Analog Interface Mode Description of the Analog Interfaces 1MA interface #1 MA interface #20x21 Configuration of the 4-20 mA Interface0x01 PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 0x001000 Reading the Selected Physical Unit for Analog InterfaceDefining the Measurement Values for 4, 12 and 20 mA Output PH ARC sensor 4-20mA inte rface linear Defining a Constant Current Output for Testing 0x010001 Reading the Internally Measured Output Current 99186 99742Measurement Definition of Measurement Channels and Physical UnitsSpecial Primary Measurement Channel 1 pH Description of PMC1Selecting the Physical Unit for PMC1 0x2010000x1000 02503 0x00 Reading the measurement value of PMC10x200000 175.9922 0x00 954.6541 Primary Measurement Channel 6 Temperature Description of PMC6Selecting the Physical Unit for PMC6 0x06Input of an Externally Measured Temperature Definition of the Measurement Status for PMC1 / PMC6Reading the measurement value of PMC6 0x04 24.35834 0x00 130Secondary Measurement Channels Description of SMCReading the measurement value of SMC 0x8000 247.56Configuration of the Measurement Available Parameters2 PA9 Moving Average Effect of Moving Average n=1 vs n=16Description of PA9 Moving Average Selecting the Physical Unit and Writing the Value for PA9Reading all Values for PA9 Description of PA12 Moving Average R 3 PA12 Moving Average RSelecting the Physical Unit and Writing the Value for PA12 Reading all Values for PA12 Calibration Available Calibration PointsDefinitions of Calibration Points Calibration Points 1 and 2 Standard CalibrationCP1 CP2Calibration Point 6 Product Calibration CP6Calibration Procedure Calibration at CP1 and CP2 Standard CalibrationPH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011 Choose one of the calibration points CP1 or CP2 0x00000000 0x00001000 100 Calibration at CP6 Product Calibration200 Definition of the commands for product calibration Product calibration Initial measurementProduct calibration Cancel Product calibration AssignmentProduct calibration Restore standard calibration Product calibration Restore product calibrationReading the Calibration Status Reading the Calibration Status of CP1 and CP20x00000180 0x00001000 0x00000080 0x00001000Reading the Calibration Status of CP6 Product Calibration 0x01000000 0x00001000 0x08000000 0x000010000x0A000000 0x00001000 0x14000000 0x000010000x10000000 0x00001000 Currently active Calibration Parameters part 0x00000004 24.35184 23.780x00000004 24.37691 16.45 0x00000004 29.93368 379.5167003401 179.927 297.1378 006804 099747 296.690110469 298.3302 607782 59.47631 298.15Special Commands for Calibration with Visical Calibration Standards Available Sets of Calibration StandardsLSB 0x00000001 Set 0x0000003FDefinitions for Individual Sets of Calibration Standards 1800x0FFF0FFF 0x02480FFF 0x05540FFFDIN Nist StandardMettler Toledo Radiometer Temperature Ranges Sensor StatusOperating Hours and Counters 168.3667History of Warnings Currently Active WarningsDefinition of Warnings Currently Active Errors ErrorsHistory of Errors Definition of Errors SIP Reading Definition of SIP and CIPCIP Reading the Sensor’s Quality Indicator 120 130Sensor Identification and Information General InformationSensor Identification EPHFI0100001001 Free User Memory SpaceFreeuserspace System Commands Recall Sensor’s Factory SettingsHello World 732255PH ARC Sensors Modbus RTU Programmer’s Manual EPHUM011