Appendix B Verification and Calibration

Test Record Form – Agilent N5746A and N5766A

Agilent N5746A and N5766A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

 

Minimum Voltage Vout

Both

0 mV

__________

+100 mV

Measurement Readback

Both

Vout 40 mV

__________

Vout + 40 mV

High Voltage Vout

Both

39.96 V

__________

40.04 V

Measurement Readback

Both

Vout 80 mV

__________

Vout + 80 mV

CV Load Effect

Both

6 mV

__________

+ 6 mV

CV Source Effect

Both

6 mV

__________

+ 6 mV

CV Ripple and Noise

 

 

 

 

 

peak-to-peak

Both

N/A

__________

60 mV

rms

Both

N/A

__________

8 mV

Transient Response

 

 

 

 

 

Voltage @ 1ms

Both

200 mV

__________

+ 200 mV

 

 

 

 

 

 

Current Programming & Readback

 

 

 

 

 

Minimum Current Iout

N5746A

0 mA

__________

+ 96 mA

 

N5766A

0 mA

__________

+ 191 mA

Measurement Readback

N5746A

Iout 57 mA

__________

Iout + 57 mA

 

N5766A

Iout 114 mA

__________

Iout + 114 mA

High Current Iout

N5746A

18.962 A

__________

19.038 A

 

N5766A

37.924 A

__________

38.076 A

Measurement Readback

N5746A

Iout 76 mA

__________

Iout + 76 mA

 

N5766A

Iout 152 mA

__________

Iout + 152 mA

CC Load Effect

N5746A

8.8 mA

__________

+ 8.8 mA

 

N5766A

12.6 mA

__________

+ 12.6 mA

CC Source Effect

N5746A

3.9 mA

__________

+ 3.9 mA

 

N5766A

5.8 mA

__________

+ 5.8 mA

 

 

 

 

 

 

 

 

Test Description

 

N5746A Setting

N5766A Setting

Voltage Programming & Readback, Min Voltage

80 mV, 76 mA

80 mV, 152 mA

Voltage Programming & Readback, High Voltage

40 V, 76 mA

 

 

40 V, 152 mA

CV Load Effect, Source Effect, Ripple and Noise

40 V, 19 A

 

 

40 V, 38 A

Transient Response

 

40 V, from 1.9 A to 17.1 A

40 V, from 3.8 A to 34.2 A

Current Programming & Readback, Min Current

76 mA, 40 V

 

 

152 mA, 40 V

Current Programming & Readback, High Current

19 A, 40 V

 

 

38 A, 40 V

CC Load Effect, Source Effect

 

19 A, 40 V

 

 

38 A, 40 V

 

 

 

 

 

 

 

 

 

 

 

 

Load Requirements

 

N5746A

 

 

N5766A

Current shunt

 

0.01100 A

 

 

0.01100 A

Agilent N3300 Electronic load modules

 

2 – N3305A

 

 

3 – N3305A

Fixed Resistor for CV Ripple and Noise

 

2.111kW

 

 

1.052kW

106

Series N5700 User’s Guide

Page 106
Image 106
Agilent Technologies 5749A, 5752A, 5744A, 5743A, 5748A, 5741A, 5751A, 5742A, 5745A Test Record Form Agilent N5746A and N5766A

5751A, 5744A, 5743A, 5750A, 5749A specifications

Agilent Technologies has been a leader in electronic measurement and analysis, with a suite of powerful instruments that cater to various applications in research, development, and manufacturing. Among these, the Agilent 5746A, 5745A, 5741A, 5748A, and 5742A models stand out for their advanced capabilities and features.

The Agilent 5746A is a high-performance signal analyzer that excels in time and frequency domain analysis. This device offers a wide frequency range and exceptional dynamic range, making it suitable for both RF and microwave applications. Its advanced digital signal processing allows for high-resolution spectrum analysis and real-time signal processing, which is essential for engineers and researchers working on complex signals.

The Agilent 5745A is known for its versatility as a multifunctional signal generator. It combines the functionalities of a vector signal generator with the precision of a traditional oscillator. This model provides high-quality waveform generation for a variety of modulation schemes, making it ideal for testing communication systems and conducting research in advanced modulation techniques.

The Agilent 5741A is a precision power meter that offers accurate measurements of RF power in a compact format. It is equipped with state-of-the-art sensing technology that ensures reliable and repeatable measurements. The 5741A is particularly useful for monitoring power levels in transmitters and receivers, and its intuitive interface makes it user-friendly for both novices and experienced professionals.

For those focused on harmonic and intermodulation distortion analysis, the Agilent 5748A provides exceptional performance. This model allows users to analyze distortion products in various devices with precision, helping in the design and testing of amplifiers and other RF components. Its comprehensive sweep functionality enables users to observe the behavior of devices across a wide frequency range.

Finally, the Agilent 5742A complements the range with its dedicated spectrum analyzing capabilities. Offering high sensitivity and ability to capture short-lived events, this model is optimal for monitoring transient signals. The 5742A is frequently used in the telecommunications sector to ensure signal integrity and compliance with regulatory standards.

Together, these Agilent Technologies instruments provide a comprehensive toolkit for professionals in electronics and telecommunications, enabling detailed analysis, accurate measurements, and advanced signal generation. Their integration of modern technologies ensures that users can tackle the complexities of today’s electronic environments with confidence and precision.