Verification and Calibration Appendix B

Test Record Form – Agilent N5747A and N5767A

Agilent N5747A and N5767A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

 

Minimum Voltage Vout

Both

0 mV

__________

+ 150 mV

Measurement Readback

Both

Vout 60 mV

__________

Vout + 60 mV

High Voltage Vout

Both

59.94 V

__________

60.06 V

Measurement Readback

Both

Vout 120 mV

__________

Vout + 120 mV

CV Load Effect

Both

8 mV

__________

+ 8 mV

CV Source Effect

Both

8 mV

__________

+ 8 mV

CV Ripple and Noise

 

 

 

 

 

peak-to-peak

Both

N/A

__________

60 mV

rms

Both

N/A

__________

8 mV

Transient Response

 

 

 

 

 

Voltage @ 1ms

Both

300 mV

__________

+ 300 mV

 

 

 

 

 

 

Current Programming & Readback

 

 

 

 

 

Minimum Current Iout

N5747A

0 mA

__________

+ 63 mA

 

N5767A

0 mA

__________

+126 mA

Measurement Readback

N5747A

Iout 37.5 mA

__________

Iout + 37.5 mA

 

N5767A

Iout 75 mA

__________

Iout + 75 mA

High Current Iout

N5747A

12.475 A

__________

12.525 A

 

N5767A

24.95 A

__________

25.05 A

Measurement Readback

N5747A

Iout 50 mA

__________

Iout + 50 mA

 

N5767A

Iout 100 mA

__________

Iout + 100 mA

CC Load Effect

N5747A

7.5 mA

__________

+ 7.5 mA

 

N5767A

10 mA

__________

+ 10 mA

CC Source Effect

N5747A

3.25 mA

__________

+ 3.25 mA

 

N5767A

4.5 mA

__________

+ 4.5 mA

 

 

 

 

 

 

 

 

Test Description

 

N5747A Setting

N5767A Setting

Voltage Programming & Readback, Min Voltage

120 mV, 50 mA

120 mV, 100 mA

Voltage Programming & Readback, High Voltage

60 V, 50 mA

 

 

60 V, 100 mA

CV Load Effect, Source Effect, Ripple and Noise

60 V, 12.5 A

 

 

60 V, 25 A

Transient Response

 

60 V, from 1.25 A to 11.25 A

60 V, from 2.5 A to 22.5 A

Current Programming & Readback, Min Current

50 mA, 60 V

 

 

100 mA, 60 V

Current Programming & Readback, High Current

12.5 A, 60 V

 

 

25 A, 60 V

CC Load Effect, Source Effect

 

12.5 A, 60 V

 

 

25 A, 60 V

 

 

 

 

 

 

 

 

 

 

 

 

Load Requirements

 

N5747A

 

 

N5767A

Current shunt

 

0.115 A

 

 

0.01100 A

Agilent N3300 Electronic load modules

 

2 – N3305A

 

 

3 – N3305A

Fixed Resistor for CV Ripple and Noise

 

4.81kW

 

 

2.42kW

Series N5700 User’s Guide

107

Page 107
Image 107
Agilent Technologies 5750A, 5752A, 5744A, 5743A, 5748A, 5741A, 5746A, 5751A, 5742A Test Record Form Agilent N5747A and N5767A

5751A, 5744A, 5743A, 5750A, 5749A specifications

Agilent Technologies has been a leader in electronic measurement and analysis, with a suite of powerful instruments that cater to various applications in research, development, and manufacturing. Among these, the Agilent 5746A, 5745A, 5741A, 5748A, and 5742A models stand out for their advanced capabilities and features.

The Agilent 5746A is a high-performance signal analyzer that excels in time and frequency domain analysis. This device offers a wide frequency range and exceptional dynamic range, making it suitable for both RF and microwave applications. Its advanced digital signal processing allows for high-resolution spectrum analysis and real-time signal processing, which is essential for engineers and researchers working on complex signals.

The Agilent 5745A is known for its versatility as a multifunctional signal generator. It combines the functionalities of a vector signal generator with the precision of a traditional oscillator. This model provides high-quality waveform generation for a variety of modulation schemes, making it ideal for testing communication systems and conducting research in advanced modulation techniques.

The Agilent 5741A is a precision power meter that offers accurate measurements of RF power in a compact format. It is equipped with state-of-the-art sensing technology that ensures reliable and repeatable measurements. The 5741A is particularly useful for monitoring power levels in transmitters and receivers, and its intuitive interface makes it user-friendly for both novices and experienced professionals.

For those focused on harmonic and intermodulation distortion analysis, the Agilent 5748A provides exceptional performance. This model allows users to analyze distortion products in various devices with precision, helping in the design and testing of amplifiers and other RF components. Its comprehensive sweep functionality enables users to observe the behavior of devices across a wide frequency range.

Finally, the Agilent 5742A complements the range with its dedicated spectrum analyzing capabilities. Offering high sensitivity and ability to capture short-lived events, this model is optimal for monitoring transient signals. The 5742A is frequently used in the telecommunications sector to ensure signal integrity and compliance with regulatory standards.

Together, these Agilent Technologies instruments provide a comprehensive toolkit for professionals in electronics and telecommunications, enabling detailed analysis, accurate measurements, and advanced signal generation. Their integration of modern technologies ensures that users can tackle the complexities of today’s electronic environments with confidence and precision.