Operating the Power Supply Locally 3
UVL Check
•Press the OVP/UVL button twice so that the DC AMPS display indicates UUL. The DC VOLTS display shows the UVL level.
•Use the voltage knob and set the UVL level of the unit to 50% of its
•Wait a few seconds until the DC VOLTS display returns to show the output voltage.
•Use the voltage knob and lower the output voltage of the unit until it approaches the UVL setting. Check to make sure that the output voltage cannot be set lower than the UVL setting.
•Press the OVP/UVL button twice. Rotate the voltage knob and reset the UVL level of the unit to its minimum setting.
Constant Current Check
•Turn the POWER switch off. Wait a few seconds until the AC indicator on the front panel goes out.
•Use a heavy wire and short the +V and
•Turn the POWER switch on.
•Turn the output on by pressing the OUT ON button. The green OUT ON indicator should be illuminated. The green CC indicator should be also illuminated.
•Rotate the current knob while observing the DC AMPS display. The output current should vary while the knob is turned. The current range is from zero to the maximum rated output for the power supply model.
OCP Check
•Rotate the current knob and set the current limit of the unit to about 10% of its
•Press the OCP/488 button. This should trip the OCP protection. The OCP indicator should be illuminated, the DC VOLTS display should indicate OCP, and the Alarm indicator should be blinking.
•Press the OCP/488 button again to cancel OCP protection. The DC VOLTS display should indicate OFF because the OCP protection is latched.
•Press the OUT ON button to reset the OCP protection. The output should return to its previous setting.
•Turn the POWER switch off.
•Remove the short from the +V and