Appendix B Verification and Calibration

Test Record Form – Agilent N5750A and N5770A

Agilent N5750A and N5770A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

 

Minimum Voltage Vout

Both

0 mV

__________

+ 375 mV

Measurement Readback

Both

Vout 150 mV

__________

Vout + 150 mV

High Voltage Vout

Both

149.85 V

__________

150.15 V

Measurement Readback

Both

Vout 300 mV

__________

Vout + 300 mV

CV Load Effect

Both

17 mV

__________

+ 17 mV

CV Source Effect

Both

17 mV

__________

+ 17 mV

CV Ripple and Noise

 

 

 

 

 

peak-to-peak

Both

N/A

__________

100 mV

rms

Both

N/A

__________

12 mV

Transient Response

 

 

 

 

 

Voltage @ 2ms

Both

750 mV

__________

+ 750 mV

 

 

 

 

 

 

Current Programming & Readback

 

 

 

 

 

Minimum Current Iout

N5750A

0 mA

__________

+ 26 mA

 

N5770A

0 mA

__________

+ 51 mA

Measurement Readback

N5750A

Iout 15 mA

__________

Iout + 15 mA

 

N5770A

Iout 30 mA

__________

Iout + 30 mA

High Current Iout

N5750A

4.99 A

__________

5.01 A

 

N5770A

9.98 A

__________

10.02 A

Measurement Readback

N5750A

Iout 20 mA

__________

Iout + 20 mA

 

N5770A

Iout 40 mA

__________

Iout + 40 mA

CC Load Effect

N5750A

6 mA

__________

+ 6 mA

 

N5770A

7 mA

__________

+ 7 mA

CC Source Effect

N5750A

2.5 mA

__________

+ 2.5 mA

 

N5770A

3 mA

__________

+ 3 mA

 

 

 

 

 

 

 

 

Test Description

 

N5750A Setting

N5770A Setting

Voltage Programming & Readback, Min Voltage

300 mV, 20 mA

300 mV, 40 mA

Voltage Programming & Readback, High Voltage

150 V, 20 mA

 

 

150 V, 40 mA

CV Load Effect, Source Effect, Ripple and Noise

150 V, 5 A

 

 

150 V, 10 A

Transient Response

 

150 V, from 0.5 A to 4.5 A

150 V, from 1 A to 9 A

Current Programming & Readback, Min Current

20 mA, 150 V

 

 

40 mA, 150 V

Current Programming & Readback, High Current

5 A, 150 V

 

 

10 A, 150 V

CC Load Effect, Source Effect

 

5 A, 150 V

 

 

10 A, 150 V

 

 

 

 

 

 

 

 

 

 

 

 

Load Requirements

 

N5750A

 

 

N5770A

Current shunt

 

0.115 A

 

 

0.115 A

Agilent N3300 Electronic load modules

 

2 – N3305A

 

 

3 – N3305A

Fixed Resistor for CV Ripple and Noise

 

30.01kW

 

 

15.02kW

110

Series N5700 User’s Guide

Page 110
Image 110
Agilent Technologies 5743A, 5752A, 5744A, 5748A, 5741A, 5746A, 5751A, 5742A, 5745A Test Record Form Agilent N5750A and N5770A

5751A, 5744A, 5743A, 5750A, 5749A specifications

Agilent Technologies has been a leader in electronic measurement and analysis, with a suite of powerful instruments that cater to various applications in research, development, and manufacturing. Among these, the Agilent 5746A, 5745A, 5741A, 5748A, and 5742A models stand out for their advanced capabilities and features.

The Agilent 5746A is a high-performance signal analyzer that excels in time and frequency domain analysis. This device offers a wide frequency range and exceptional dynamic range, making it suitable for both RF and microwave applications. Its advanced digital signal processing allows for high-resolution spectrum analysis and real-time signal processing, which is essential for engineers and researchers working on complex signals.

The Agilent 5745A is known for its versatility as a multifunctional signal generator. It combines the functionalities of a vector signal generator with the precision of a traditional oscillator. This model provides high-quality waveform generation for a variety of modulation schemes, making it ideal for testing communication systems and conducting research in advanced modulation techniques.

The Agilent 5741A is a precision power meter that offers accurate measurements of RF power in a compact format. It is equipped with state-of-the-art sensing technology that ensures reliable and repeatable measurements. The 5741A is particularly useful for monitoring power levels in transmitters and receivers, and its intuitive interface makes it user-friendly for both novices and experienced professionals.

For those focused on harmonic and intermodulation distortion analysis, the Agilent 5748A provides exceptional performance. This model allows users to analyze distortion products in various devices with precision, helping in the design and testing of amplifiers and other RF components. Its comprehensive sweep functionality enables users to observe the behavior of devices across a wide frequency range.

Finally, the Agilent 5742A complements the range with its dedicated spectrum analyzing capabilities. Offering high sensitivity and ability to capture short-lived events, this model is optimal for monitoring transient signals. The 5742A is frequently used in the telecommunications sector to ensure signal integrity and compliance with regulatory standards.

Together, these Agilent Technologies instruments provide a comprehensive toolkit for professionals in electronics and telecommunications, enabling detailed analysis, accurate measurements, and advanced signal generation. Their integration of modern technologies ensures that users can tackle the complexities of today’s electronic environments with confidence and precision.