Agilent Technologies 5748A, 5752A, 5744A, 5743A, 5741A Test Record Form Agilent N5751A and N5771A

Models: 5751A 5744A 5743A 5750A 5749A 5752A 5747A 5742A 5748A 5741A 5745A 5746A

1 132
Download 132 pages 2.14 Kb
Page 111
Image 111

Verification and Calibration Appendix B

Test Record Form – Agilent N5751A and N5771A

Agilent N5751A and N5771A

Report No _______________

Date __________________

Description

Model

Minimum Specs.

Results

Maximum Specs.

Voltage Programming & Readback

 

 

 

 

 

Minimum Voltage Vout

Both

0 mV

__________

+ 750 mV

Measurement Readback

Both

Vout 300 mV

__________

Vout + 300 mV

High Voltage Vout

Both

299.7 V

__________

300.3 V

Measurement Readback

Both

Vout 600 mV

__________

Vout + 600 mV

CV Load Effect

Both

32 mV

__________

+ 32 mV

CV Source Effect

Both

32 mV

__________

+ 32 mV

CV Ripple and Noise

 

 

 

 

 

peak-to-peak

Both

N/A

__________

150 mV

rms

Both

N/A

__________

20 mV

Transient Response

 

 

 

 

 

Voltage @ 2ms

Both

1 V

__________

+ 1 V

 

 

 

 

 

 

Current Programming & Readback

 

 

 

 

 

Minimum Current Iout

N5751A

0 mA

__________

+ 13 mA

 

N5771A

0 mA

__________

+ 26 mA

Measurement Readback

N5751A

Iout 7.5 mA

__________

Iout + 7.5 mA

 

N5771A

Iout 15 mA

__________

Iout + 15 mA

High Current Iout

N5751A

2.495 A

__________

2.505 A

 

N5771A

4.99 A

__________

5.01 A

Measurement Readback

N5751A

Iout 10 mA

__________

Iout + 10 mA

 

N5771A

Iout 20 mA

__________

Iout + 20 mA

CC Load Effect

N5751A

5.5 mA

__________

+ 5.5 mA

 

N5771A

6 mA

__________

+ 6 mA

CC Source Effect

N5751A

2.25 mA

__________

+ 2.25 mA

 

N5771A

2.5 mA

__________

+ 2.5 mA

 

 

 

 

 

 

 

 

Test Description

 

N5751A Setting

N5771A Setting

Voltage Programming & Readback, Min Voltage

600 mV, 10 mA

600 mV, 20 mA

Voltage Programming & Readback, High Voltage

300 V, 10 mA

 

 

300 V, 20 mA

CV Load Effect, Source Effect, Ripple and Noise

300 V, 2.5 A

 

 

300 V, 5 A

Transient Response

 

200 V, from 0.25 A to 2.25 A

200 V, from 0.5 A to 4.5 A

Current Programming & Readback, Min Current

10 mA, 300 V

 

 

20 mA, 300 V

Current Programming & Readback, High Current

2.5 A, 300 V

 

 

5 A, 300 V

CC Load Effect, Source Effect

 

2.5 A, 300 V

 

 

5 A, 300 V

 

 

 

 

 

 

 

 

 

 

 

 

Load Requirements

 

N5751A

 

 

N5771A

Current shunt

 

0.115 A

 

 

0.115 A

Use fixed resistor instead of load modules

1201kW

 

 

602kW

Fixed Resistor for CV Ripple and Noise

 

1201kW

 

 

602kW

 

 

 

 

 

 

Series N5700 User’s Guide

111