Operating the Power Supply Remotely 4
Connecting to a Private LAN:
A private LAN is a network in which
To Network | To LAN Port |
Interface Card (NIC) |
|
CAT5 Crossover Cable | |
PC | Instrument |
1If you have not already done so, install the Agilent IO Libraries Suite from the
2Connect the instrument to the computer using a LAN crossover cable. Alternatively, connect the computer and the instrument to a standalone hub or switch using regular LAN cables.
| Make sure your computer is configured to obtain its address from DHCP and that |
NOTE | |
| NetBIOS over TCP/IP is enabled. If the computer had been connected to a site LAN, |
| it may still retain previous network settings from the site LAN. Wait one minute |
| after disconnecting it from the site LAN before connecting it to the private LAN. This |
| allows Windows to sense that it is on a different network and restart the network |
| configuration. If you are running Windows 98, you may need to manually release the |
| previous settings. |
|
|
| 3 The |
| automatically obtain an IP address from the network using a |
| DHCP server, or using AutoIP if a DHCP server is not present. |
| You can leave these settings as they are. Most Agilent products |
| and most computers will automatically choose an IP address |
| using |
| an IP address from the block 169.254.nnn. Note that this may |
| take up to one minute. |
| The front panel LAN indicator will come on when the LAN port |
| has been configured. If you are unable to communicate with the |
| instrument, check that a valid IP address has been assigned. |
| Press the front panel LAN button to view the IP address. |
| 4 Use the Connection Expert utility of the Agilent IO Libraries |
| Suite to add the N5700 power supply and verify a connection. To |
| add the instrument, you can request the Connection Expert to |
| discover the instrument. If the instrument cannot be found, you |
| can add the instrument using the instrument’s hostname. The |
| default hostname is described under “Connecting to a Site LAN”. |
| If this does not work, refer to the chapter on “Troubleshooting Guidelines” in the |
NOTE | |
| Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide. |
|
|
Series N5700 User’s Guide | 51 |