Agilent Technologies 5752A, 5744A, 5743A, 5748A, 5741A, 5746A, 5751A, 5742A, 5745A, 5747A Command Errors

Models: 5751A 5744A 5743A 5750A 5749A 5752A 5747A 5742A 5748A 5741A 5745A 5746A

1 132
Download 132 pages 2.14 Kb
Page 120
Image 120

Appendix C Service

Command Errors (continued)

141 Invalid character data

Either the character data element contains an invalid character, or the element is not valid.

144 Character data too long

The character data element contains more than 12 characters.

148 Character data not allowed

A discrete parameter was received, but a string or numeric parameter was expected.

150 String data error Generic string data error

151 Invalid string data

An invalid character string was received. Check that the string is enclosed in quotation marks.

158 String data not allowed

A character string was received, but is not allowed for this command.

160 Block data error Generic block data error

161 Invalid block data

The number of data bytes sent does not match the number of bytes specified in the header.

168 Block data not allowed

Data was sent in arbitrary block format but is not allowed for this command.

170 Expression error Generic expression error

171 Invalid expression data

The expression data element was invalid.

178 Expression data not allowed

Expression data element was sent but is not allowed for this command.

Execution Errors (these errors set Standard Event Status register bit #4)

200 Execution error

Generic syntax error

220 Parameter error

A data element related error occurred.

221 Settings conflict

A data element could not be executed because of the present instrument state.

222 Data out of range

A data element could not be executed because the value was outside the valid range.

223 Too much data

A data element was received that contains more data than the instrument can handle.

224 Illegal parameter value

An exact value was expected but not received.

225 Out of memory

The device has insufficient memory to perform the requested operation.

226 Lists not same length

One or more lists are not the same length.

230 Data corrupt or stale

Possible invalid data. A new reading was started but not completed.