5 Language Reference

*IDN?

This query requests the power supply to identify itself. It returns a string of four fields separated by commas.

Agilent Technologies

Manufacturer

xxxxxA

Model number followed by a letter suffix

0

Zero or serial number if available

<A.xx.xx>,<A.xx.xx>

Firmware revision, power supply revision

*OPT?

This query requests the unit to identify any installed options. A 0 indicates no options are installed.

*RCL <state>

 

This command restores the power supply to a state that was

 

previously stored in memory locations 0 through 15 with the *SAV

 

command. Note that you can only recall a state from a location that

 

contains a previously-stored state.

 

All saved instrument states are lost when the unit is turned off.

NOTE

 

 

*RST

This command resets the power supply to a factory-defined state. This state is defined as follows. Note that *RST also forces an ABORt command. The *RST settings are as follows:

CAL:STAT

Off

[SOUR:]CURR:PROT:STAT

Off

INIT:CONT

Off

[SOUR:]VOLT

0

OUTP

Off

[SOUR:]VOLT:LIM

0

[SOUR:]CURR

0

[SOUR:]VOLT:TRIG

0

[SOUR:]CURR:TRIG

0

[SOUR:]VOLT:PROT

MAXimum

*SAV <state>

 

This command stores the present state of the power supply to

 

memory locations 0 through 15.

NOTE

All saved instrument states are lost when the unit is turned off.

*TST?

This query always returns a zero.

78

Series N5700 User’s Guide