Table 1-1. Performance Specifications (continued)
Parameter
Line Regulation (change in output voltage or current for any line change within ratings)
Voltage:
Current:
Transient Response Time (for output voltage to recover to within 100 mV of its previous level following a step change in load current that is up to 10% of the power module rated output current.
AC Input Ratings (selectable via module switch- see Chapter 2)
Nominal line voltage (Vac) 115 V switch setting 230 V switch setting
Frequency range:
Output Terminal Isolation (maximum, from output to the mainframe chassis ground):
Agilent Model
66101A | 66102A | 66103A | 66104A | 66105A | 66106A |
0.5 mV | 0.5 mV | 1 mV | 2 mV | 3 mV | 5 mV |
0.75 mA | 0.5. mA | 0.3 mA | 0.1 mA | 50 μA | 30 μA |
< 1 ms
87- 132 Vac
174- 250 Vac
47 - 63 Hz
±240 Vdc
Table 1-2. Supplemental Characteristics
Parameter
Output Programming Range (Maximum programmable values):
Agilent Model
66101A | 66102A | 66103A | 66104A | 66105A | 66106A |
Voltage:
Current:
OV Protection:
Average Programming Resolution
Voltage:
Current:
OV Protection:
OVP Accuracy (@ calibration temperature ±5° C):*
8.190 V
16.380 A
9.6V
2.4mV
4.6mA
50 mV
250 mV
20.475 V
7.678 A
24.0V
5.9mV
2.3mA
120 mV
500 mV
35.831 V
4.607 A
42.0V
10.4mV 1.4 mA 200 mV 800 mV
61.425 V
2.559 A
72.0V
18.0mV
0.75mA
375 mV
1 V
122.85V 1.280 A 144.0 V
36.0mV
0.39mA
750 mV
1.5V
204.75V 0.768 A 240.0 V
60.0mV
0.23mA
1.25V 2.5 V
*Factory calibration temp = 25° C
Readback Resolution Voltage:
Current:
0.305 mV | 0.763 mV | 1.335 mV | 2.289 mV | 4.577 mV | 7.629 mV |
0.587 mA | 0.293 mA | 0.169 mA | 0.099 mA | 0.049 mA | 0.030 mA |
Drift (Temperature Stability) (following a
Voltage: | 0.02% + |
Current: | 0.02% + |
0.4mV 8 mA
1mV
5mA
2mV 2.5 mA
3mV 1.3 mA
6mV
0.63 mA
10mV 0.38 mA