CURR:PROT:STAT OFF OUTP:PROT:CLE

Power Module Turn-On State

When you turn on the mainframe, the power module goes through a self-test. All front panel readout digits and annunciators turn on briefly. The panel then remains in the state shown in Table 3-1A.

Note

The figure assumes the power module still retains its factory default turn-on state. You can change this

 

state if desired (see "Chapter 4 - Basic Power Module Commands").

 

 

Checking Basic Module Functions

The power module is designed for remote programming over the GPIB. It also can be locally programmed from the optional Agilent 66001A MPS Keyboard. Table 3-2 gives the SCPI commands for the remotely programmed tests. You can send these same commands from the keyboard (see "Appendix B").

Command

Table 3-2. Testing Basic Functions with SCPI Commands

Action

 

 

Display Response

 

VOLTS

AMPS

 

CV

CC

Addr*

Dis

Prot

Output Voltage (Disconnect Load)

VOLT 5.1

Set output voltage to 5.1 V

OUTP ON

Enable the output

Overvoltage Protection

VOLT:PROT 4.9

Activate OVP circuit (by

 

setting protection voltage

 

below output voltage)

VOLT:PROT MAX

Clear OVP circuit

OUTP:PROT:CLE

Clear OVP circuit

Save and Recall Functions

*SAV 5

Save present state to

 

location 5

VOLT 3.55

Set output voltage to 3.55 V

OUTP OFF

Turn off the output

*SAV 6

Save present state to

 

location 6

*RCL 5

Restore state 5

*RCL 6

Restore state 6

Output Current (With output disabled, short the output terminals with a wire size sufficient to carry the maximum current of the module.)

0.00

5.10

0.00

5.10

5.10

5.10

3.55

0.00

0.00

5.10

0.00

0.00

0.00

0.00

0.00

0.00

0.00

0.00

0.00

0.00

0.00

0.00

Off

On

Off

Off On

On

On

Off

Off

On Off

Off

Off

Off

Off Off

Off

Off

Off

Off

Off Off

On

On

On

On On

On

On

On

On

On On

On

Off

Off

Off Off

Off

Off

On

On

Off On

Off

Off

On

Off Off

Off

Off

Off

Off

Off Off

CURR 3.1

Set output current to 3.1

 

amps

OUTP ON

Enable the output

Overcurrent Protection

CURR:PROT:STAT

Activate OCP circuit (by

ON

enabling OCP when the

 

output is shorted)

 

Clear OCP circuit

 

 

0.00

0.00

0.00

0.00

0.00

3.10

0.00

0.00

Off

Off

Off

Off

Off

On

Off

On

On

On

On

On

On

Off

Off

On

Off

Off

On

Off

Clear OCP circuit

0.00

3.10

Off

On

*Addr will not light if the module is controlled from the keyboard.

On

Off

Off

28 Turn-On Checkout

Page 28
Image 28