Agilent Technologies 66001A manual Power Module Line Fuse and Switches

Models: 66001A

1 55
Download 55 pages 3.42 Kb
Page 18
Image 18

Figure 2-1. Power Module Line Fuse and Switches

 

Figure 2-2. Power Module Configuration Switch

 

Table 2-1. Settings for Power Module Configuration Switch

 

 

 

 

 

Required Switch Setting

 

 

Desired Function

 

1

2

3

4

5

6

7*

8*

RI disabled

 

0

0

 

 

 

 

1

1

RI unlatched (live)

 

0

1

 

 

 

 

1

1

RI latching †

 

1

1

 

 

 

 

1

1

Display disabled

 

 

 

0

 

 

 

1

1

Display enabled †

 

 

 

I

 

 

 

1

1

Inhibit calibration

 

 

 

 

0

0

 

1

1

Factory calibration

 

 

 

 

0

1

 

1

1

Normal calibration †

 

 

 

 

1

1

 

1

1

Power on state is *RCL 0

 

 

 

 

 

0

1

1

Power on state is *RST†

 

 

 

 

 

 

1

1

1

†= factory default setting

"0" Off or open

"1" On or closed.

 

 

 

 

*Positions 7 and 8 are for service functions. For normal operation, they must be set to 1.

18 Installation

Page 18
Image 18
Agilent Technologies 66001A manual Power Module Line Fuse and Switches