Point (GP).

 

12.

The CP should be connected to the GP as shown in Figures 40 through 43 (unless one load is

72

 

already grounded), making certain there is only one conductive path between these two points.

 

13.

Connections between the power supply sensing and output terminals should be removed and

74

 

using shielded two-wire cable, the power supply sensing terminals should be connected to the

 

 

DC Distribution Terminals as shown in Figure 49.

 

14.

One end of the shield should be connected to the CP and the other end should be left

75

 

unconnected.

 

15.

The possibility of an open remote sensing path, which might occur on a long-term or transient

75

 

basis, should be avoided.

 

16.

The minimum wire size for the load current leads (from the power supply output terminals to

77

 

the DT's) should be determined for remote sensing.

 

17.

Check for the possibility of power supply oscillation when connected in the system for remote

77

 

sensing.

 

18.

Check for proper current limiting operation while the power supply is connected in the system

78

 

for remote sensing.

 

 

Load Connections for Two or More Power Supplies

 

19.

There must be only one point of connection between the dc outputs of any two power supplies

79

 

in the multiple power supply system—this point must be designated as one of the two DT's for

 

 

both power supplies.

 

20.

One of the (N + 1) DT's determined in accordance with the preceding rule is designated as the

79

 

CP for the system.

 

21.

There must be only one GP per multiple power supply system.

79

22.

Connect the System CP to the System GP (unless one load is already grounded), making

79

 

certain there is only one conductive path between these two points for the entire system.

 

AC POWER INPUT CONNECTIONS

The ac, acc and third wire safety ground continuity should be retained without accidental interchange from ac power outlet to the power supply input terminals.

Accidental interchanging of ac and safety ground leads may result in the power supply chassis being elevated to an ac potential equal to the line input voltage. This could result in a potentially lethal shock hazard, if the chassis is not grounded; or blown fuses, if the chassis is grounded.

If ac and acc are accidentally interchanged, the power supply switches and fuses are thereby placed in series with the ground side of the power line instead of the hot side-if the power supply ac line switch is turned off or the fuse opens, the hot side of the power line will be connected to exposed components within the power supply.

Accidental interchanging of acc and ground leads places the chassis at the acc potential giving rise to circulating ground currents flowing through the power supply chassis and other associated ground return paths- the result is often excessive power supply output ripple and malfunction of associated instrument.

60

Page 59
Image 59
Agilent Technologies 90B manual Load Connections for Two or More Power Supplies, AC Power Input Connections