Chapter 3 Calibration Procedures
Constant Current (CC) Verifications
7Program the output current to the full rated value (7.0A) by sending the commands:
CURR 7.0
8Divide the voltage drop (DVM reading) across the current monitoring resistor
(RM) by its resistance to convert to amps and record this value (IO). This value should be within the limit of (7A ± 24mA).
9Readback the output current over the remote interface by sending the command:
MEAS:CURR?
10Record the value displayed on the controller. This value should be within the limit (IO ± 15.5mA).
CC Load Regulation
This test measures the change in output current resulting from a change in the load from
1Turn off the power supply and connect the output to tested as shown in Figure
2Turn on the power supply. Select the 15V/7A range, enable the output, and set the display to the limit mode. When the display is in the limit mode, program the output voltage to the full rated value (15.0V) and the output current to the full rated value (7.0A).
3Operate the electronic load in constant voltage mode and set its voltage to 15.0V. Check that the CC annunciator is on. If it is not, adjust the load so that the output voltage drops slightly. Record the current reading by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor.
4Operate the electronic load in short (input short) mode. Record the current reading again by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor. The difference between the current readings in step (3) and (4) is the load regulation current. The difference of the readings should be within the limit of 0.95mA.