Chapter 3 Calibration Procedures

Automating Calibration Procedures

Automating Calibration Procedures

You can automate the complete verification procedures outlined in this chapter if you have access to programmable test equipment. You can program the instrument configurations specified for each test over the remote interface. You can then enter readback verification data into a test program and compare the results to the appropriate test limit values.

You can also enter calibration constants from the remote interface. Remote operation is similar to the local front-panel procedure. You can use a computer to perform the adjustment by first selecting the required setup. The calibration value is sent to the power supply and then the calibration is initiated over the remote interface. The power supply must be unsecured prior to initiating the calibration procedure. An Agilent BASIC program for calibration over the GPIB interface is listed at the end of this chapter.

For further details on programming the power supply, see chapters 3 and 4 in the Agilent E3632A User's Guide.

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, use the accuracy requirements shown to select substitute calibration standards.

Table 3-1 Recommended Test Equipment

Instrument

Requirements

Recommended

Use

Model

 

 

 

 

 

 

 

GPIB Controller

Full GPIB capabilities

Agilent 82341C

Programming and

Interface card

readback accuracy

 

 

 

 

 

 

 

100 MHz with 20MHz

 

Display transient response

Oscilloscope

Agilent 54602B

and ripple & noise

bandwidth

 

 

waveform

 

 

 

 

 

 

 

RMS Voltmeter

20 MHz

 

Measure rms ripple & noise

 

 

 

 

Digital Voltmeter

Resolution: 0.1 mV

Agilent 34401A

Measure dc voltages

Accuracy: 0.01%

 

 

 

 

 

 

 

 

Voltage Range: 50 Vdc

 

Measure load and line

 

Current Range: 10 Adc

 

Electronic Load

Agilent 6063B

regulations and transient

Open and Short Switches

 

Transient On/Off

 

response time.

 

 

 

 

 

 

 

Resistive Loads (RL)

2.19, 200 W

 

Measure ripple and noise

7.59, 200 W

 

 

 

 

 

Current monitoring

0.019, 0.01%

 

Constant current test setup

Resistor (Shunt)

 

 

 

 

 

 

 

 

44

Page 45
Image 45
Agilent Technologies E3632A manual Automating Calibration Procedures, Recommended Test Equipment