Chapter 3 Calibration Procedures Calibration Record for Agilent E3632A
Calibration Record for Agilent E3632A
Step | Calibration Description | Measurement | Supply being | |
Mode (DVM) | Adjusted | |||
|
| |||
|
|
|
| |
1 | Unsecure the power supply (see page 59). |
|
| |
|
|
|
| |
2 | Turn on "CAL MODE"(simultaneously press "Calibrate" and "Power" keys) |
|
| |
until it beeps. |
|
| ||
|
|
| ||
|
|
|
| |
3 | Move down menu to "CAL SETUP 1" (press "Calibrate" key). |
| Voltage | |
| Calibration | |||
|
|
| ||
|
|
|
| |
| Calibrate DAC and select the low point for voltage calibration; "START |
|
| |
| BITCAL" appears for 3 seconds and the display counts down numbers from |
| DAC and low | |
4 | 30 to 0. Then, " V LO 0.5000 V" appears on the display(press "Calibrate" key | V | voltage point | |
| and wait about 30 seconds; then change the display to match the DVM |
| calibration | |
| reading). |
|
| |
|
|
|
| |
5 | "V MI 15.000 V" appears on the display (press "Calibrate" key; then change | V | Middle voltage | |
the display to match the DVM reading). | point calibration | |||
|
| |||
|
|
|
| |
6 | "V HI 29.500 V" appears on the display (press "Calibrate" key; then change | V | High voltage point | |
the display to match the DVM reading). | calibration | |||
|
| |||
|
|
|
| |
7 | "CAL SETUP 2" now appears on the display (press "Calibrate" key). | V | OVP calibration | |
|
|
|
| |
8 | "CAL SETUP 3" now appears on the display (press "Calibrate" key; then |
| Current | |
connect 0.019resistor across the output terminals). |
| calibration | ||
|
| |||
|
|
|
| |
9 | "I LO 0.200 A" appears on the display (press "Calibrate" key; then change | A | Low current point | |
the display to match the computed current through 0.019resistor). | calibration | |||
|
| |||
|
|
|
| |
10 | "I MI 3.500 A" appears on the display (press "Calibrate" key; then change the | A | Middle current | |
display to match the computed current through 0.019resistor). | point calibration | |||
|
| |||
|
|
|
| |
11 | "I HI 6.9000 A" appears on the display (press "Calibrate" key; then change | A | High current point | |
the display to match the computed current through 0.019resistor). | calibration | |||
|
| |||
|
|
|
| |
12 | "CAL SETUP 4" now appears on the display (press "Calibrate" key). | A | OCP calibration | |
|
|
|
| |
13 | Press "Calibrate" key, then press "Power" switch. |
| Exit CAL MODE | |
|
|
|
|