Chapter 3 Calibration Procedures Calibration Record for Agilent E3632A

Calibration Record for Agilent E3632A

Step

Calibration Description

Measurement

Supply being

Mode (DVM)

Adjusted

 

 

 

 

 

 

1

Unsecure the power supply (see page 59).

 

 

 

 

 

 

2

Turn on "CAL MODE"(simultaneously press "Calibrate" and "Power" keys)

 

 

until it beeps.

 

 

 

 

 

 

 

 

 

3

Move down menu to "CAL SETUP 1" (press "Calibrate" key).

 

Voltage

 

Calibration

 

 

 

 

 

 

 

 

Calibrate DAC and select the low point for voltage calibration; "START

 

 

 

BITCAL" appears for 3 seconds and the display counts down numbers from

 

DAC and low

4

30 to 0. Then, " V LO 0.5000 V" appears on the display(press "Calibrate" key

V

voltage point

 

and wait about 30 seconds; then change the display to match the DVM

 

calibration

 

reading).

 

 

 

 

 

 

5

"V MI 15.000 V" appears on the display (press "Calibrate" key; then change

V

Middle voltage

the display to match the DVM reading).

point calibration

 

 

 

 

 

 

6

"V HI 29.500 V" appears on the display (press "Calibrate" key; then change

V

High voltage point

the display to match the DVM reading).

calibration

 

 

 

 

 

 

7

"CAL SETUP 2" now appears on the display (press "Calibrate" key).

V

OVP calibration

 

 

 

 

8

"CAL SETUP 3" now appears on the display (press "Calibrate" key; then

 

Current

connect 0.019resistor across the output terminals).

 

calibration

 

 

 

 

 

 

9

"I LO 0.200 A" appears on the display (press "Calibrate" key; then change

A

Low current point

the display to match the computed current through 0.019resistor).

calibration

 

 

 

 

 

 

10

"I MI 3.500 A" appears on the display (press "Calibrate" key; then change the

A

Middle current

display to match the computed current through 0.019resistor).

point calibration

 

 

 

 

 

 

11

"I HI 6.9000 A" appears on the display (press "Calibrate" key; then change

A

High current point

the display to match the computed current through 0.019resistor).

calibration

 

 

 

 

 

 

12

"CAL SETUP 4" now appears on the display (press "Calibrate" key).

A

OCP calibration

 

 

 

 

13

Press "Calibrate" key, then press "Power" switch.

 

Exit CAL MODE

 

 

 

 