Calibrate

Calibrate

Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

v lo +0.5000 v

Then, the display shows the low voltage calibration point.

6Read the DVM and change the low voltage value on the display to match the measured voltage.

For example, if the DVM reading is 0.4500 V, adjust the voltage to 0.4500 V using the knob and resolution selection keys.

v lo +0.4500 v

7Pressing the "Calibrate" key saves the change and selects the middle voltage calibration point.

v Mi +15.000v

If the entered number is within an acceptable range, an "ENTERED" message appears for one second. If the entered number is not correct, an "MIN VALUE" "MAX VALUE" message appears for one second and the display shows the low voltage calibration point again. The display now shows the middle voltage calibration point.

8Read the DVM and change the middle voltage value on the display to match the measured voltage.

For example, if the DVM reads 14.995 V, adjust the voltage to 14.995 V using the knob and arrow keys.

v Mi +14.995v

9Pressing the "Calibrate" key saves the change and selects the high voltage calibration point.

v hi 29.500 v

64

Page 65
Image 65
Agilent Technologies E3632A manual Lo +0.5000