Store

Chapter 2 Quick Start

To Store and Recall the Instrument State

To Store and Recall the Instrument State

You can store up to three different operating states in non-volatile memory. This also enables you to recall the entire instrument with just a few key presses from the front panel.

The memory locations are supplied with the reset state from the factory for front panel operation. Refer to the description of *RST command, starting on page 94 in the User’s Guide for more information.

The following steps show you how to store and recall an operating state.

1Set up the power supply for the desired operating state.

The storage feature "remembers" the output range selection, the limit value settings of voltage and current, output on/off state , OVP and OCP on/off state, and OVP and OCP trip levels.

2Turn on the storage mode.

Three memory locations (numbered 1, 2, and 3) are available to store the operating states. The operating states are stored in non-volatile memory and are remembered when being recalled.

Store 1

This message appears on the display for approximately 3 seconds.

3Store the operating state in memory location "3".

Turn the knob to the right to specify the memory location 3.

Store 3

To cancel the store operation, let the display time-out after about 3 seconds or press any other function key except the Store key. The power supply returns to the normal operating mode and to the function pressed.

32

Page 33
Image 33
Agilent Technologies E3632A manual To Store and Recall the Instrument State