Calibrate

Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

16Read the DVM and change the middle current value on the display to match the computer current (DVM reading ¸ by shunt resistance).

For example, if the computered value is 3.499 A, adjust the current to 3.499 A using the knob and arrow keys.

Notice that you should wait for the DVM reading to stabilize for accurate calibration.

i mi 3.499 a

17Pressing the "Calibrate'' key saves the change and selects the high current calibration point.

i hi 6.900 a

If the entered number is within an acceptable range, an "ENTERED" message appears for one second. If the entered number is not correct, an "MIN VALUE" or "MAX VALUE" message appears for one second and the display shows the middle current calibration point again. The display now shows the high current calibration point

18Read the DVM and change the high current value on the display to match the computed current (DVM reading ¸ by shunt resistance).

For example, if the computed value is 6.899 A, adjust the current to 6.899 A using the knob and arrow keys.

Notice that you should wait for the DVM reading to stabilize for accurate calibration.

i hi 6.899 a

67

3

Page 68
Image 68
Agilent Technologies E3632A manual Mi 3.499 a