Chapter 2 Quick Start

To Program Overvoltage Protection

5Exit the OVP menu.

changed

The "CHANGED" message is displayed for a second to show that the new OVP trip level is now in effect. If the OVP settings are not changed, "NO CHANGE" will be displayed. The power supply will exit the OVP menu and the display will return to the meter mode. Check that the OVP annunciator turns on.

Checking OVP Operation

To check OVP operation, raise the output voltage to near the trip point. Then very gradually increase the output by turning the knob until the OVP circuit trips. This will cause the power supply output to drop to near zero. The OVP annunciator is blinking and the CC annunicator turns on. The "OVP TRIPPED" message also appears on the display.

Clearing the Overvoltage Condition

2

Over Voltage

When the OVP circuit trips (the "OVP TRIPPED" message is shown on the display), the OVP annunicator blinks. When it was caused by external voltage source such as a battery, disconnect it first. The following steps show how to clear the overvoltage conditions and get back to normal mode operation. In the following steps, the display will go back to "OVP TRIPPED" if you let the display time out after about several seconds.

or

Display Limit

Over Voltage

1Readjust the OVP trip level or the output voltage level.

Lower the output voltage level below the OVP trip point after pressing the

Display Limit or raise the OVP trip level by using the knob after pressing

the Over Voltage key.

2 Move to the clear mode.

ovp on

You will see the above message after pressing the Over Voltage key. If you changed the output voltage level, press the Over Voltage key twice. Turn the knob to the right until the "OVP CLEAR" appears on the display.

35

Page 36
Image 36
Agilent Technologies E3632A Checking OVP Operation, Clearing the Overvoltage Condition, Move to the clear mode Ovp on