Chapter 3 Calibration Procedures Performance Test Record for Agilent E3632A

Performance Test Record for Agilent E3632A

CV Performance Test Record

Test Description

Actual

Specifications

 

 

Result

Upper Limit

Lower Limit

 

 

 

 

 

 

 

CV Programming Accuracy @ 0 volts

 

+0.0100 V

-0.0100 V

(DVM reading)

 

 

 

 

 

 

 

 

CV Readback Accuracy @ 0 volts

 

DVM + 0.0050 V

DVM - 0.0050 V

 

 

 

 

CV Programming Accuracy @ Full Scale

 

+30.025V

29.9750 V

(DVM reading)

 

 

 

 

 

 

 

 

CV Readback Accuracy @ Full Scale

 

DVM + 0.0200 V

DVM - 0.0200 V

 

 

 

 

CV Load Regulation

 

Maximum change: <5 mV

 

 

 

CV Line Regulation

 

Maximum change: <5 mV

 

 

 

CV Ripple/Noise

 

<2 mV p-p, 0.35 mV rms

 

 

 

 

Load Transient Response Time

 

<50 msec

 

 

 

 

 

CC Performance Test Record

Test Description

Actual

Specifications

 

 

Result

Upper Limit

Lower Limit

 

 

 

 

 

 

 

CC Programming Accuracy @ 0 amps (IO)

 

+0.0100 A

-0.0100 A

CC Readback Accuracy @ 0 amps

 

IO + 0.0050 A

IO - 0.0050 A

CC Programming Accuracy @ Full Scale (IO)

 

7.0240 A

6.9760 A

CC Readback Accuracy @ Full Scale

 

IO + 0.0155 A

IO - 0.0155 A

CC Load Regulation

 

Maximum change: <0.95 mA

 

 

 

CC Line Regulation

 

Maximum change: <0.95 mA

 

 

 

 

CC Ripple/Noise

 

<2 mA

 

 

 

 

 

Common Mode Current Noise

 

<1.5 mA rms

 

 

 

 

 