Chapter 3 Calibration Procedures

Calibration Program

continued

1860

!

 

 

1870

CLEAR SCREEN

 

1880

IF Error$="+0,""No error""" THEN

 

1890

PRINT "Current calibration completed for Power Supply "

1900

ELSE

 

 

1910

PRINT "ERROR:";Error$;"Current not Calibrated"

1920

BEEP

 

 

1930

GOTO

2250

 

1940

END IF

 

 

1950

CLEAR SCREEN

 

1960

PRINT TABXY(10,5),"CURRENT CALIBRATION COMPLETE"

1970

PRINT TABXY(10,7),"BEGIN OCP CALIBRATION"

1980

WAIT 4

 

 

1990

CLEAR SCREEN

 

2000

OUTPUT

@Pwrsupply;"OUTP ON"

! Turn on Power Supply output

2010

OUTPUT

@Pwrsupply;"CAL:CURR:PROT"

! Perform OCP calibration

2020

WAIT 9

 

! Allow OCP calibration to finish

2030

OUTPUT

@Pwrsupply;"OUTP OFF"

! Turn off Power Supply output

2040

OUTPUT

@Pwrsupply;"SYST:ERR?"

 

2050

ENTER @Pwrsupply;Error$

 

2060

!

 

 

2070

! Check to see if there is an error.

If there is an error,

2080

! display the error and exit the program.

2090

!

 

 

2100

CLEAR SCREEN

 

2110

IF Error$="+0,""No error""" THEN

 

2120

PRINT "OCP calibration completed for Power Supply "

2130

ELSE

 

 

2140

PRINT "ERROR:";Error$;"OCP not Calibrated"

2150

BEEP

 

 

2160

GOTO

2250

 

2170

END IF

 

 

2180

CLEAR SCREEN

 

2190

PRINT TABXY(10,5),"CURRENT/OCP CALIBRATION COMPLETE"

2200

!

 

 

2210

! Create a time stamp and output to power supply

2220

!

 

 

2230

Cal_msg$="Last Calibrated "&DATE$(TIMEDATE)&" "&TIME$(TIMEDATE)

2240

OUTPUT

@Pwrsupply;"CAL:STR """;Cal_msg$;""""

2250

OUTPUT

@Pwrsupply;"CAL:SEC:STAT ON, ";Sec_code$

2260

OUTPUT

@Pwrsupply;"VOLT:PROT:STAT ON"

 

2270

OUTPUT

@Pwrsupply;"CURR:PROT:STAT ON"

 

2280

DISP "Calibration terminated. "

 

2290

END

 

 

End of Program