Display Annunciators

Adrs

Power supply is addressed to listen or talk over a remote interface.

Rmt

Power supply is in remote interface mode.

15V

Shows the 15V/7A range is selected.

30V

Shows the 30V/4A range is selected.

OVP

The overvoltage protection function is enabled when the annunciator

 

turns on or the overvoltage protection circuit has caused the power

 

supply to shutdown when the annunciator blinks.

OCP

The overcurrent protection function is enabled when the annunciator

 

turns on or the overcurrent protection circuit has caused the power

 

supply to shutdown when the annunciator blinks.

CAL

The power supply is in calibration mode.

Limit

The display shows the limit values of voltage and current.

ERROR

Hardware or remote interface command errors are detected and the error

 

bit has not been cleared.

OFF

The output of the power supply is disabled (See page 52 for more

 

information).

Unreg

The output of the power supply is unregulated (output is neither CV

 

nor CC).

CV

The power supply is in constant voltage mode.

CCThe power supply is in constant current mode.

To review the display annunciators, hold down turn on the power supply.

Display Limit

key as you

5

Page 6
Image 6
Agilent Technologies E3632A manual Display Annunciators, Ovp