Calibrating Reflective Media

1Lift each of the printheads for this calibration routine in order to conserve ribbon. Disengage the drive roller by turning the Drive Roll Lift Lever clockwise.

2Position the reflective mark so that the red light from the reflective sensor passes through the mark. The light will typically pass through most media allowing you to view the mark from above. On more opaque media, you may have to view the mark from the bottom.

3Engage the drive roller by turning the Drive Roll Lift Lever counterclockwise.

4Press the PRINT/PAUSE key until the printer display shows Pause on the display.

5Press PREV or NEXT key until the Calibrate menu is displayed.

6Press the + or - key until Reflective is displayed.

Pause: Calibrate

Reflective *

7Press the SELECT key. The display will show the following prompt:

Place Mark @ Sensor

Then Press Select

8Press the SELECT key again. The display will show the following prompt:

Place Media @ Sensor

Then Press Select

9Reposition the media by pressing and holding the FORM FEED/ADVANCE key until the red light no longer shines through the mark but through the media.